Design and Fabrication of Kidney Phantoms for Internal Radiation Dosimetry Using 3D Printing Technology.
Journal of Nuclear Medicine 2016 December
Currently, the validation of multimodal quantitative imaging and absorbed dose measurements is impeded by the lack of suitable, commercially available anthropomorphic phantoms of variable sizes and shapes. To demonstrate the potential of 3-dimensional (3D) printing techniques for quantitative SPECT/CT imaging, a set of kidney dosimetry phantoms and their spherical counterparts was designed and manufactured with a fused-deposition-modeling 3D printer. Nuclide-dependent SPECT/CT calibration factors were determined to assess the accuracy of quantitative imaging for internal renal dosimetry.
METHODS: A set of 4 single-compartment kidney phantoms with filling volumes between 8 and 123 mL was designed on the basis of the outer kidney dimensions provided by MIRD pamphlet 19. After the phantoms had been printed, SPECT/CT acquisitions of 3 radionuclides (99m Tc, 177 Lu, and 131 I) were obtained and calibration constants determined for each radionuclide-volume combination. A set of additionally manufactured spheres matching the kidney volumes was also examined to assess the influence of phantom shape and size on the calibration constants.
RESULTS: A set of refillable, waterproof, and chemically stable kidneys and spheres was successfully manufactured. Average calibration factors for 99m Tc, 177 Lu, and 131 I were obtained in a large source measured in air. For the largest phantom (122.9 mL), the volumes of interest had to be enlarged by 1.2 mm for 99m Tc, 2.5 mm for 177 Lu, and 4.9 mm for 131 I in all directions to obtain calibration factors comparable to the reference. Although partial-volume effects were observed for decreasing phantom volumes (percentage difference of up to 9.8% for the smallest volume [8.6 mL]), the difference between corresponding sphere-kidney pairs was small (<1.1% for all volumes).
CONCLUSION: 3D printing is a promising prototyping technique for geometry-specific calibration of SPECT/CT systems. Although the underlying radionuclide and the related collimator have a major influence on the calibration, no relevant differences between kidney-shaped and spherically shaped uniform-activity phantoms were observed. With comparably low costs and submillimeter resolution, 3D printing techniques hold the potential for manufacturing individualized anthropomorphic phantoms in many clinical applications in nuclear medicine.
METHODS: A set of 4 single-compartment kidney phantoms with filling volumes between 8 and 123 mL was designed on the basis of the outer kidney dimensions provided by MIRD pamphlet 19. After the phantoms had been printed, SPECT/CT acquisitions of 3 radionuclides (99m Tc, 177 Lu, and 131 I) were obtained and calibration constants determined for each radionuclide-volume combination. A set of additionally manufactured spheres matching the kidney volumes was also examined to assess the influence of phantom shape and size on the calibration constants.
RESULTS: A set of refillable, waterproof, and chemically stable kidneys and spheres was successfully manufactured. Average calibration factors for 99m Tc, 177 Lu, and 131 I were obtained in a large source measured in air. For the largest phantom (122.9 mL), the volumes of interest had to be enlarged by 1.2 mm for 99m Tc, 2.5 mm for 177 Lu, and 4.9 mm for 131 I in all directions to obtain calibration factors comparable to the reference. Although partial-volume effects were observed for decreasing phantom volumes (percentage difference of up to 9.8% for the smallest volume [8.6 mL]), the difference between corresponding sphere-kidney pairs was small (<1.1% for all volumes).
CONCLUSION: 3D printing is a promising prototyping technique for geometry-specific calibration of SPECT/CT systems. Although the underlying radionuclide and the related collimator have a major influence on the calibration, no relevant differences between kidney-shaped and spherically shaped uniform-activity phantoms were observed. With comparably low costs and submillimeter resolution, 3D printing techniques hold the potential for manufacturing individualized anthropomorphic phantoms in many clinical applications in nuclear medicine.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app