BACKGROUND: Measurement of postoperative serum thyroglobulin (Tg) is important for detecting persistent or recurrent differentiated thyroid cancer. We evaluated the analytic performance of the DxI 800 assay (Beckman Coulter, USA) for serum Tg and anti-thyroglobulin antibodies (TgAbs) in comparison with that of the GAMMA-10 assay (Shinjin Medics Inc., Korea) for serum Tg and RIA-MAT 280 assay (Stratec, Germany) for TgAb.
METHODS: We prospectively collected blood samples from 99 patients thyroidectomized for thyroid cancer. The functional sensitivity was investigated in standards and human serum. Precision and linearity were evaluated according to the guidelines of the Clinical and Laboratory Standards Institute. The correlation between the two assays was assessed in samples with different Tg ranges.
RESULTS: The functional sensitivity of the DxI 800 assay for serum Tg was between 0.0313 and 0.0625 ng/mL. The total CV was 3.9-5.6% for serum Tg and 5.3-6.9% for serum TgAb. The coefficient of determination (R²) was 1.0 and 0.99 for serum Tg and TgAb, respectively. The cut-offs for serum TgAb were 4.0 IU/mL (DxI 800) and 60.0 IU/mL (RIA-MAT 280), and the overall agreement was 68.7%. The correlation between the two assays was excellent; the correlation coefficient was 0.99 and 0.88 for serum Tg and TgAb, respectively.
CONCLUSIONS: The DxI 800 is a sensitive assay for serum Tg and TgAb, and the results correlated well with those from the immunoradiometric assays (IRMA). This assay has several advantages over the IRMA and could be considered an alternative test for Tg measurement.
METHODS: We prospectively collected blood samples from 99 patients thyroidectomized for thyroid cancer. The functional sensitivity was investigated in standards and human serum. Precision and linearity were evaluated according to the guidelines of the Clinical and Laboratory Standards Institute. The correlation between the two assays was assessed in samples with different Tg ranges.
RESULTS: The functional sensitivity of the DxI 800 assay for serum Tg was between 0.0313 and 0.0625 ng/mL. The total CV was 3.9-5.6% for serum Tg and 5.3-6.9% for serum TgAb. The coefficient of determination (R²) was 1.0 and 0.99 for serum Tg and TgAb, respectively. The cut-offs for serum TgAb were 4.0 IU/mL (DxI 800) and 60.0 IU/mL (RIA-MAT 280), and the overall agreement was 68.7%. The correlation between the two assays was excellent; the correlation coefficient was 0.99 and 0.88 for serum Tg and TgAb, respectively.
CONCLUSIONS: The DxI 800 is a sensitive assay for serum Tg and TgAb, and the results correlated well with those from the immunoradiometric assays (IRMA). This assay has several advantages over the IRMA and could be considered an alternative test for Tg measurement.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app