We have located links that may give you full text access.
Comparison of I-gel with Classic Laryngeal Mask Airway Regarding the Ease of Use and Clinical Performance.
Turkish Journal of Anaesthesiology and Reanimation 2015 October
OBJECTIVE: I-gel is a new supraglottic airway device without an inflatable cuff. We aimed to compare I-gel and the classic laryngeal mask airway (LMA) regarding the ease of use and clinical performance in Turkish population.
METHODS: Fifty American Society of Anesthesiologists (ASA) I-II patients were randomly allocated into two groups: Group I-gel and Group LMA. Insertion time and success in first attempt were recorded. Peak, plato and mean airway pressures, EtCO2, airway compliance and leak volume were periodically recorded during the operation. The presence of blood on device removal and postoperative sore throat were also assessed.
RESULTS: The device insertion time in Group I-gel was shorter than that in Group LMA (21.00±4.15 vs. 30.40±12.17 s, p=0.001). The success rate in first attempt, peak, plato and mean airway pressures, EtCO2 and airway compliance did not differ between the groups. The leak volume was lower in Group I-gel 5 and 45 min after insertion (p=0.041 and p=0.027). The presence of blood on device removal and postoperative sore throat were similar in both groups.
CONCLUSION: I-gel may be a more advantageous supraglottic airway device compared with LMA.
METHODS: Fifty American Society of Anesthesiologists (ASA) I-II patients were randomly allocated into two groups: Group I-gel and Group LMA. Insertion time and success in first attempt were recorded. Peak, plato and mean airway pressures, EtCO2, airway compliance and leak volume were periodically recorded during the operation. The presence of blood on device removal and postoperative sore throat were also assessed.
RESULTS: The device insertion time in Group I-gel was shorter than that in Group LMA (21.00±4.15 vs. 30.40±12.17 s, p=0.001). The success rate in first attempt, peak, plato and mean airway pressures, EtCO2 and airway compliance did not differ between the groups. The leak volume was lower in Group I-gel 5 and 45 min after insertion (p=0.041 and p=0.027). The presence of blood on device removal and postoperative sore throat were similar in both groups.
CONCLUSION: I-gel may be a more advantageous supraglottic airway device compared with LMA.
Full text links
Related Resources
Trending Papers
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app