Evaluation of the Sebia Capillarys 3 Tera and the Bio-Rad D-100 Systems for the Measurement of Hemoglobin A1c.
American Journal of Clinical Pathology 2016 July
OBJECTIVES: We evaluated the Bio-Rad (Irvine, CA) D-100 and the Sebia (Lisses, France) Capillarys 3 Tera for the measurement of hemoglobin A1c (HbA1c) in venous blood samples.
METHODS: Whole-blood samples and control material were analyzed with the D-100 and Capillarys 3 Tera and compared with our routine method, HLC-723G7 (Tosoh, Tokyo, Japan). An evaluation protocol to test precision, trueness, linearity, carryover, and selectivity was set up according to Clinical and Laboratory Standards Institute guidelines. The results were presented in National Glycohemoglobin Standardization Program and International Federation of Clinical Chemistry (IFCC) units.
RESULTS: Both systems showed excellent precision (total coefficients of variation <2%, IFCC) and bias (<0.3% or 3 mmol/mol). Linearity was demonstrated for HbA1c values from 3.8% (18 mmol/mol) to 18.5% (179 mmol/mol). Results were correlated with the routine method using Bland-Altman analysis, showing a mean difference of 0.33% or 3.6 mmol/mol for the D-100 and of 0.25% or 2.6 mmol/mol for the Capillarys 3 Tera vs HLC-723G7. None of the automated instruments were prone to interferences by labile HbA1c (≤10 g/L glucose), carbamylated hemoglobin (≤0.5 mmol/L potassium cyanate), hemoglobin variants, bilirubin (≤15 mg/dL), and triglycerides (≤3,360 mg/dL).
CONCLUSIONS: The Bio-Rad D-100 and the Sebia Capillarys 3 Tera instruments performed well for the determination of HbA1c in terms of quality criteria as well as for sample throughput.
METHODS: Whole-blood samples and control material were analyzed with the D-100 and Capillarys 3 Tera and compared with our routine method, HLC-723G7 (Tosoh, Tokyo, Japan). An evaluation protocol to test precision, trueness, linearity, carryover, and selectivity was set up according to Clinical and Laboratory Standards Institute guidelines. The results were presented in National Glycohemoglobin Standardization Program and International Federation of Clinical Chemistry (IFCC) units.
RESULTS: Both systems showed excellent precision (total coefficients of variation <2%, IFCC) and bias (<0.3% or 3 mmol/mol). Linearity was demonstrated for HbA1c values from 3.8% (18 mmol/mol) to 18.5% (179 mmol/mol). Results were correlated with the routine method using Bland-Altman analysis, showing a mean difference of 0.33% or 3.6 mmol/mol for the D-100 and of 0.25% or 2.6 mmol/mol for the Capillarys 3 Tera vs HLC-723G7. None of the automated instruments were prone to interferences by labile HbA1c (≤10 g/L glucose), carbamylated hemoglobin (≤0.5 mmol/L potassium cyanate), hemoglobin variants, bilirubin (≤15 mg/dL), and triglycerides (≤3,360 mg/dL).
CONCLUSIONS: The Bio-Rad D-100 and the Sebia Capillarys 3 Tera instruments performed well for the determination of HbA1c in terms of quality criteria as well as for sample throughput.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app