COMPARATIVE STUDY
JOURNAL ARTICLE
Add like
Add dislike
Add to saved papers

A new cervical artificial disc prosthesis based on physiological curvature of end plate: a finite element analysis.

STUDY DESIGN: The study aimed to build a new cervical artificial disc C3-C7 segment prosthesis, and perform a biomechanical comparison between the new prosthesis and the Prestige LP prosthesis using a three-dimensional non-linear finite element (FE) model.

PURPOSE: The study compared the biomechanical differences between the new cervical artificial disc prosthesis based on the physiological curvature of the end plate and the Prestige LP prosthesis after artificial disc replacement.

BACKGROUND CONTEXT: There has been no prior research on artificial disc prostheses based on the physiological curvature of the end plate; studies of biomechanical changes after cervical disc arthroplasty (CDR) are few.

METHODS: An FE model of the C3-C7 segments was developed and validated. A new cervical artificial disc prosthesis based on the physiological curvature of the end plate and the Prestige LP prosthesis were integrated at the C5-C6 segment into the validated FE model. All models were subjected to a follower load of 73.6 N and a 1 Nm in flexion-extension, lateral bending, and axial torsion. The segmental range of motion (ROM) and stress on the prostheses were analyzed.

RESULTS: The ROM in most segments after CDR with new cervical artificial disc prosthesis was more similar to that of the normal cervical spine than the Prestige LP prosthesis. However, there was no significant difference between the two prostheses. The stress on the new artificial disc was significantly less than that in the Prestige LP prosthesis.

CONCLUSIONS: There was no significant difference in ROM in all segments after CDR for the two prostheses. The stress on the new cervical artificial disc prosthesis based on the physiological curvature of the end plate was significantly less than that in the Prestige LP prosthesis. The new artificial disc prosthesis is feasible and effective, and can reduce the implant-bone interface stress on the end plate, which may be one of the causes of prosthesis subsidence.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app