Evaluation of a new PT-INR monitoring system in patients with the antiphospholipid syndrome.
International Journal of Laboratory Hematology 2016 October
INTRODUCTION: Patients on anticoagulant therapy with vitamin K antagonists (VKA) need frequent INR monitoring. Reliability of point-of-care (POC) devices for measuring INR needs rigorous evaluation, particularly in patient with the antiphospholipid syndrome (APS). The aim of this study was to evaluate the accuracy of the ProTime InRhythm(™) System (here called device) for INR measurement in patients with APS on VKA.
METHODS: We compared the device INR vs. the laboratory INR measurement for blood samples from 29 APS-positive and 31 APS-negative patients consecutively enrolled. APS was confirmed by positive serological and/or phospholipid-dependent coagulation tests. Chromogenic factor X assay was used to evaluate anticoagulation. Bland-Altman difference plot for paired INR (POC vs. laboratory) was used to evaluate agreement between the device and the laboratory. The device INR relationship with factor X chromogenic assay was evaluated by orthogonal regression analysis.
RESULTS: Overall, 97% of the device INR measurements were similar to laboratory INR values with an absolute difference less than 0.4 units. Correlation coefficient for the device INR vs. factor X was -0.69 (P < 0.0001, CI 95% -0.80 to -0.52).
CONCLUSIONS: The ProTime InRhythm System(™) is an accurate point-of-care device for measuring INR also in patients with and without APS.
METHODS: We compared the device INR vs. the laboratory INR measurement for blood samples from 29 APS-positive and 31 APS-negative patients consecutively enrolled. APS was confirmed by positive serological and/or phospholipid-dependent coagulation tests. Chromogenic factor X assay was used to evaluate anticoagulation. Bland-Altman difference plot for paired INR (POC vs. laboratory) was used to evaluate agreement between the device and the laboratory. The device INR relationship with factor X chromogenic assay was evaluated by orthogonal regression analysis.
RESULTS: Overall, 97% of the device INR measurements were similar to laboratory INR values with an absolute difference less than 0.4 units. Correlation coefficient for the device INR vs. factor X was -0.69 (P < 0.0001, CI 95% -0.80 to -0.52).
CONCLUSIONS: The ProTime InRhythm System(™) is an accurate point-of-care device for measuring INR also in patients with and without APS.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app