Long-term right ventricular implantable cardioverter-defibrillator lead performance in arrhythmogenic right ventricular cardiomyopathy.
BACKGROUND: Arrhythmogenic right ventricular cardiomyopathy (ARVC) is a progressive disease characterized by replacement of normal myocardium by fibrofatty tissue. The right ventricular (RV) apex is the typical target for implantable cardioverter-defibrillator (ICD) lead placement, raising concerns for suboptimal lead performance in medium- to long-term follow-up.
OBJECTIVE: The purpose of this study was to determine whether placement of ICD leads at the RV apex was associated with performance deterioration of medium-term leads in ARVC patients compared to non-ARVC patients.
METHODS: In this multicenter, retrospective, case-control study, ICD lead performance measures of R-wave, impedance, and pacing thresholds were compared at baseline and between 1-year and 5-year postimplantation follow-up using mixed-effect models adjusted for age and sex.
RESULTS: One hundred one ARVC patients (49 women, age 50.6 ± 14.5 years) were compared to 56 control patients (37 women, age 48.2 ± 14.2 years). The mean difference in R wave between years 1 and 2 was -0.85 mV (P = .16) compared to a mean difference at years 5 and 6 of -1.85 mV (P = .02). There was no difference in impedance or pacing threshold or in lead lifetime between the 2 groups over 6-year follow-up (5.91 ± 3.89 years vs 5.48 ± 3.70 years, P = .239).
CONCLUSION: In ARVC patients with ICD leads implanted in the RV apex, ventricular sensing deteriorates significantly during medium-term follow-up. Septal RV lead placement should be explored as the first choice at implantation.
OBJECTIVE: The purpose of this study was to determine whether placement of ICD leads at the RV apex was associated with performance deterioration of medium-term leads in ARVC patients compared to non-ARVC patients.
METHODS: In this multicenter, retrospective, case-control study, ICD lead performance measures of R-wave, impedance, and pacing thresholds were compared at baseline and between 1-year and 5-year postimplantation follow-up using mixed-effect models adjusted for age and sex.
RESULTS: One hundred one ARVC patients (49 women, age 50.6 ± 14.5 years) were compared to 56 control patients (37 women, age 48.2 ± 14.2 years). The mean difference in R wave between years 1 and 2 was -0.85 mV (P = .16) compared to a mean difference at years 5 and 6 of -1.85 mV (P = .02). There was no difference in impedance or pacing threshold or in lead lifetime between the 2 groups over 6-year follow-up (5.91 ± 3.89 years vs 5.48 ± 3.70 years, P = .239).
CONCLUSION: In ARVC patients with ICD leads implanted in the RV apex, ventricular sensing deteriorates significantly during medium-term follow-up. Septal RV lead placement should be explored as the first choice at implantation.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app