12-Month Coronary Angiography, Intravascular Ultrasound and Histology Evaluation of a Novel Fully Bioabsorbable Poly-L-Lactic Acid/Amorphous Calcium Phosphate Scaffolds in Porcine Coronary Arteries.

Our previous studies have confirmed the superior biocompatibility of the poly-L-lactic acid/amorphous calcium phosphate (PLLA/ACP) scaffolds (PowerScaffold) compared to PLLA scaffolds and their similar 6-month radial strength compared with TAXUS stents. In order to conduct further dynamic observations on the performance of the PowerScaffold after 12-month implantation compared with the TAXUS stents. Twenty PowerScaffold and 20 TAXUS were implanted in porcine coronary arteries. At 12-month follow-up, Quantitative Coronary Angiography showed that the stent reference vessel diameter (3.19 ± 0.25 mm vs. 2.75 ± 0.22 mm, p < 0.05), the mean lumen diameter (3.07 ± 0.22 mm vs. 2.70 ± 0.17 mm, p < 0.05) and the late lumen gain (0.45 ± 0.07 mm vs. 0.06 ± 0.06 mm, p < 0.01) were all significantly greater with the PowerScaffold than the TAXUS. As well, Intravascular Ultrasound showed the stent reference vessel area (7.74 ± 0.48 mm2 vs. 6.96 ± 0.51 mm2, p < 0.05), the mean stent area (7.49 ± 0.46 mm2 vs. 6.53 ± 0.47 mm2, p < 0.05) and the mean lumen area (7.22 ± 0.50 mm2 vs. 6.00 ± 0.48 mm2, p < 0.01) were all significantly greater with the PowerScaffold than the TAXUS. The luminal patency rate of the PowerScaffold significantly increased from 72.45 ± 6.84% at 1 month to 93.54 ± 8.15% at 12 months (p < 0.01) while the TAXUS stents were associated with a non-significant decreasing trend (89.44 ± 8.44% vs. 86.53 ± 8.22%). Pathology indicated the average thickness of the struts degraded by 14.25 ± 3.04 μm at 1 month, 23.39 ± 2.45 μm at 6 months and 35.54 ± 2.20 μm at 12 months. Immunohistochemical examination showed that the expression of inflammatory factors NF-κB gradually decreased from 1-month to 12-month (36.79 ± 4.78 vs. 5.79 ± 2.85, P < 0.01). As the late lumen gain of arteries implanted with the PowerScaffold increases over time with the growth of vessels, it effectively reverse the late vascular negative remodeling observed with the TAXUS stents, providing a better option for lumen restoration treatment in clinical practice.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app