Add like
Add dislike
Add to saved papers

The phenomenon of drop in output at larger field sizes for telecobalt units.

It is known that the output factors (OPFs) for external-beam radiotherapy units increase with field size due to increased scattered radiation from the collimator system. Saturation in the OPF value is generally reported beyond approximately 30 × 30 cm(2). For the first time, to the best of our knowledge, we report on a drop in OPF values, although marginal, measured for a telecobalt machine beyond the 38 × 38 cm(2) field size. We believe that reporting and explaining the results will lead to a better understanding of the scatter composition of the radiation from telecobalt machines. This also has the potential to impact the estimation of low dose regions in patients, in addition to being a purely scientific inquiry. We used Monte Carlo (MC) simulations to validate the measured values. The MC data showed that the decrease in OPF was due to decreased scatter from the machine head.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app