Intraocular pressure study using monitored forced-infusion system phacoemulsification technology.
PURPOSE: To optimize phacoemulsification efficiency by varying the intraocular pressure (IOP) using the Centurion Vision System, a monitored forced-infusion system.
SETTING: John A. Moran Eye Center Laboratories, University of Utah, Salt Lake City, Utah, USA.
DESIGN: Laboratory investigation.
METHODS: Porcine lenses were soaked in formalin and divided into 2.0 mm cubes; 0.9 mm balanced tips were used. Intraocular pressure levels were tested from 30 to 110 mm Hg at 10 mm Hg intervals. Vacuum was set at 500 mm Hg, aspiration at 35 mL/min, and torsional power at 60%. Efficiency (time to lens removal) and chatter (number of lens fragment repulsions from the tip) were measured.
RESULTS: There was a linear relationship between efficiency and IOP. Increasing IOP led to an increase in efficiency (ie, to a decrease in time necessary to remove the lens fragment) (R(2) = .5769, P = .0176). Chatter correlated with IOP in the range of 30 to 50 mm Hg (R(2) = .4506, P = .0448) and was minimal at higher IOPs.
CONCLUSIONS: Increasing IOP in the monitored forced-infusion system improved efficiency. Chatter was also decreased or eliminated with increasing IOP. Additional studies will be required to understand the reasons for these findings.
FINANCIAL DISCLOSURE: None of the authors has a financial or proprietary interest in any material or method mentioned.
SETTING: John A. Moran Eye Center Laboratories, University of Utah, Salt Lake City, Utah, USA.
DESIGN: Laboratory investigation.
METHODS: Porcine lenses were soaked in formalin and divided into 2.0 mm cubes; 0.9 mm balanced tips were used. Intraocular pressure levels were tested from 30 to 110 mm Hg at 10 mm Hg intervals. Vacuum was set at 500 mm Hg, aspiration at 35 mL/min, and torsional power at 60%. Efficiency (time to lens removal) and chatter (number of lens fragment repulsions from the tip) were measured.
RESULTS: There was a linear relationship between efficiency and IOP. Increasing IOP led to an increase in efficiency (ie, to a decrease in time necessary to remove the lens fragment) (R(2) = .5769, P = .0176). Chatter correlated with IOP in the range of 30 to 50 mm Hg (R(2) = .4506, P = .0448) and was minimal at higher IOPs.
CONCLUSIONS: Increasing IOP in the monitored forced-infusion system improved efficiency. Chatter was also decreased or eliminated with increasing IOP. Additional studies will be required to understand the reasons for these findings.
FINANCIAL DISCLOSURE: None of the authors has a financial or proprietary interest in any material or method mentioned.
Full text links
Related Resources
Trending Papers
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app