Clinical safety and performance of VIVIA: a novel home hemodialysis system.
Nephrology, Dialysis, Transplantation 2017 April 2
Background: The VIVIA Hemodialysis System (Baxter Healthcare Corporation, Deerfield, IL, USA) was designed for patient use at home to reduce the burden of treatment and improve patient safety. It has unique features including extended use of the dialyzer and blood set through in situ hot-water disinfection between treatments; generation of on-line infusible-quality dialysate for automated priming, rinseback and hemodynamic support during hypotension and a fully integrated access disconnect sensor.
Methods: The safety and performance of VIVIA were assessed in two clinical studies. A first-in-man study was a prospective, single-arm study that involved 22 prevalent hemodialysis (HD) patients who were treated for ∼4 h, four times a week, for 10 weeks. A second clinical study was a prospective, single-arm study (6-8 h of dialysis treatment at night three times a week) that involved 17 prevalent patients treated for 6 weeks.
Results: There were 1114 treatments from the two studies (first-in-man study, 816; extended duration study, 298). Adverse events (AEs) were similar in the two studies to those expected for prevalent HD patients. No deaths and no device-related serious AEs occurred. Adequacy of dialysis ( Kt / V ) urea in both clinical trials was well above the clinical guidelines. VIVIA performed ultrafiltration accurately as prescribed in the two studies. The majority of patients achieved 10 or more uses of the dialyzer. Endotoxin levels and bacterial dialysate sampling met infusible-quality dialysate standards.
Conclusion: These results confirm the safety and expected performance of VIVIA.
Methods: The safety and performance of VIVIA were assessed in two clinical studies. A first-in-man study was a prospective, single-arm study that involved 22 prevalent hemodialysis (HD) patients who were treated for ∼4 h, four times a week, for 10 weeks. A second clinical study was a prospective, single-arm study (6-8 h of dialysis treatment at night three times a week) that involved 17 prevalent patients treated for 6 weeks.
Results: There were 1114 treatments from the two studies (first-in-man study, 816; extended duration study, 298). Adverse events (AEs) were similar in the two studies to those expected for prevalent HD patients. No deaths and no device-related serious AEs occurred. Adequacy of dialysis ( Kt / V ) urea in both clinical trials was well above the clinical guidelines. VIVIA performed ultrafiltration accurately as prescribed in the two studies. The majority of patients achieved 10 or more uses of the dialyzer. Endotoxin levels and bacterial dialysate sampling met infusible-quality dialysate standards.
Conclusion: These results confirm the safety and expected performance of VIVIA.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app