Assessment of low contrast detection in CT using model observers: Developing a clinically-relevant tool for characterising adaptive statistical and model-based iterative reconstruction.
Zeitschrift Für Medizinische Physik 2017 June
PURPOSE: This study aims to assess CT image quality in a way that would meet specific requirements of clinical practice. Physics metrics like Fourier transform derived metrics were traditionally employed for that. However, assessment methods through a detection task have also developed quite extensively lately, and we chose here to rely on this modality for image quality assessment. Our goal was to develop a tool adapted for a fast and reliable CT image quality assessment in order to pave the way for new CT benchmarking techniques in a clinical context. Additionally, we also used this method to estimate the benefits brought by some IR algorithms.
MATERIALS AND METHODS: A modified QRM chest phantom containing spheres of 5 and 8mm at contrast levels of 10 and 20HU at 120kVp was used. Images of the phantom were acquired at CTDIvol of 0.8, 3.6, 8.2 and 14.5mGy, before being reconstructed using FBP, ASIR 40 and MBIR on a GE HD 750 CT scanner. They were then assessed by eight human observers undergoing a 4-AFC test. After that, these data were compared with the results obtained from two different model observers (NPWE and CHO with DDoG channels). The study investigated the effects of the acquisition conditions as well as reconstruction methods.
RESULTS: NPWE and CHO models both gave coherent results and approximated human observer results well. Moreover, the reconstruction technique used to retrieve the images had a clear impact on the PC values. Both models suggest that switching from FBP to ASIR 40 and particularly to MBIR produces an increase of the low contrast detection, provided a minimum level of exposure is reached.
CONCLUSION: Our work shows that both CHO with DDoG channels and NPWE models both approximate the trend of humans performing a detection task. Both models also suggest that the use of MBIR goes along with an increase of the PCs, indicating that further dose reduction is still possible when using those techniques. Eventually, the CHO model associated to the protocol we described in this study happened to be an efficient way to assess CT images in a clinical environment. In the future, this simple method could represent a sound basis to benchmark clinical practice and CT devices.
MATERIALS AND METHODS: A modified QRM chest phantom containing spheres of 5 and 8mm at contrast levels of 10 and 20HU at 120kVp was used. Images of the phantom were acquired at CTDIvol of 0.8, 3.6, 8.2 and 14.5mGy, before being reconstructed using FBP, ASIR 40 and MBIR on a GE HD 750 CT scanner. They were then assessed by eight human observers undergoing a 4-AFC test. After that, these data were compared with the results obtained from two different model observers (NPWE and CHO with DDoG channels). The study investigated the effects of the acquisition conditions as well as reconstruction methods.
RESULTS: NPWE and CHO models both gave coherent results and approximated human observer results well. Moreover, the reconstruction technique used to retrieve the images had a clear impact on the PC values. Both models suggest that switching from FBP to ASIR 40 and particularly to MBIR produces an increase of the low contrast detection, provided a minimum level of exposure is reached.
CONCLUSION: Our work shows that both CHO with DDoG channels and NPWE models both approximate the trend of humans performing a detection task. Both models also suggest that the use of MBIR goes along with an increase of the PCs, indicating that further dose reduction is still possible when using those techniques. Eventually, the CHO model associated to the protocol we described in this study happened to be an efficient way to assess CT images in a clinical environment. In the future, this simple method could represent a sound basis to benchmark clinical practice and CT devices.
Full text links
Related Resources
Trending Papers
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app