Surveillance of guideline practices for duodenoscope and linear echoendoscope reprocessing in a large healthcare system.

BACKGROUND AND AIMS: To assess the adequacy of currently recommended duodenoscope and linear echoendoscope (DLE) automatic endoscope reprocessing (AER) and high-level disinfection (HLD), we collected daily post-reprocessing surveillance cultures of 106 DLEs in 21 Providence and Affiliate Hospitals.

METHODS: Daily qualitative surveillance of dried, post-HLD DLEs was conducted for a minimum of 30 days at each facility. Positivity rates for any microbial growth and growth of high-concern pathogens were reported. Potential effects of DLE manufacturer, age, and AER processor on culture-positivity rate were assessed.

RESULTS: Microbial growth was recovered from 201 of 4032 specimens (5%) or 189 of 2238 encounters (8.4%), including 23 specimens (.6%) or 21 encounters (.9%) for a high-concern pathogen. Wide variations in culture-positivity rate were observed across facilities. No striking difference in culture-positivity rate was seen among 8 DLE models, 3 DLE manufacturers, DLE age, manual or bedside cleanser, or automatic flushing system use. However, there was suggestive evidence that Custom Ultrasonics AER (Warminster, Pa, USA) had a lower culture-positivity rate than Medivators AER (Cantel Medical Corp., Little Falls, NJ, USA) for high-concern pathogen growth (0/1079 vs 21/2735 specimens or 0/547 vs 20/1582 encounters). Two endoscopes grew intestinal flora on several occasions despite multiple HLD. No multidrug-resistant organism was detected.

CONCLUSIONS: In this multicenter DLE surveillance study, microbial growth was recovered in 5.0% of specimens (8.4% of encounters), with most being environmental microbes. Enteric bacterial flora was recovered in .6% of specimens (.9% of encounters), despite compliance with 2014 U.S. guidelines and manufacturers' recommendations for cleaning and HLD process. The observed better performance of Custom Ultrasonics AER deserves further investigation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app