Case Reports
Journal Article
Add like
Add dislike
Add to saved papers

Perclose Closure Device breakage and embolization during deployment followed by retrieval with snare.

Indian Heart Journal 2015 December
Perclose Closure Device (Abbott-Vascular) was attempted for femoral artery (FA) access site closure by an experienced operator. The device was felt to snap as it was pushed through the previously scarred skin and subcutaneous tissue. Perclose had broken into two pieces with the proximal portion outside the body and the distal portion embolizing to lie in the FA to distal aorta. The left FA was accessed through 8F sheath and 4-8mm Basket snare passed through a 6F multipurpose guiding catheter. The tip of the broken Perclose was caught with the snare and pulled out. The Perclose has a distal flexible sheath that is overmolded directly onto the proximal rigid guide. This junction of sheath and guide may be a weak point. Accidental pressure applied at this weak point in any Perclose device during deployment can cause breakage. Awareness of this rare complication is important for safety.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app