Cobalt serum levels differ in well functioning Birmingham resurfacing and Birmingham modular THA.
INTRODUCTION: Metal-on-metal (MoM) bearings are known to release metal ions secondary to wear and corrosion. This may cause local reactions (adverse soft tissue reactions and osteolysis) and systemic effects. Little is known about the exact pattern and the differences between large head MoM total hip replacements (THA) and resurfacings (HR).
QUESTIONS: (1) Is there a difference in metal ion concentrations between HR and MoM-THR using the same bearing design (Birmingham Hip Resurfacing System, Smith & Nephew, Inc. Memphis, TN, USA)? (2) Are metal ion levels changing over time in MoM-THA or HR? (3) Do acetabular inclination angle and femoral component size influence cobalt and chromium levels? Is there a correlation between clinical outcome and metal ion levels?
MATERIALS AND METHODS: A retrospective analysis was conducted in 77 well functioning unilateral Birmingham HR and 42 well functioning unilateral modular Birmingham MoM-THA (Smith & Nephew, Inc. Memphis, TN, USA) operated on between 2007 and 2012. Blood samples were taken at a minimum of 13 months and subsequent during annual follow-ups.
RESULTS: (1) Cobalt levels were significantly higher in MoM-THA compared to HR (p < 0.001). There was no significant difference in chromium levels (p = 0.313). (2) Cobalt is increasing over time in MoM-THA (p = 0.030) whereas metal ions remain stable in HR. (3) Metal ion levels were not affected by acetabular inclination angle and femoral component size in MoM-THA. Chromium levels correlate with the femoral component size (r = -0.240; p = 0.037), the UCLA activity score (r = -0.344; p = 0.003) and the VAS (r = 0.263; p = 0.38) in HR.
CONCLUSION: Considering that HR and MoM-THA used the same MoM bearing design, increased cobalt levels may be related to trunnion wear or corrosion. Elevated cobalt levels should raise concern for corrosion related failure in MoM-THA.
QUESTIONS: (1) Is there a difference in metal ion concentrations between HR and MoM-THR using the same bearing design (Birmingham Hip Resurfacing System, Smith & Nephew, Inc. Memphis, TN, USA)? (2) Are metal ion levels changing over time in MoM-THA or HR? (3) Do acetabular inclination angle and femoral component size influence cobalt and chromium levels? Is there a correlation between clinical outcome and metal ion levels?
MATERIALS AND METHODS: A retrospective analysis was conducted in 77 well functioning unilateral Birmingham HR and 42 well functioning unilateral modular Birmingham MoM-THA (Smith & Nephew, Inc. Memphis, TN, USA) operated on between 2007 and 2012. Blood samples were taken at a minimum of 13 months and subsequent during annual follow-ups.
RESULTS: (1) Cobalt levels were significantly higher in MoM-THA compared to HR (p < 0.001). There was no significant difference in chromium levels (p = 0.313). (2) Cobalt is increasing over time in MoM-THA (p = 0.030) whereas metal ions remain stable in HR. (3) Metal ion levels were not affected by acetabular inclination angle and femoral component size in MoM-THA. Chromium levels correlate with the femoral component size (r = -0.240; p = 0.037), the UCLA activity score (r = -0.344; p = 0.003) and the VAS (r = 0.263; p = 0.38) in HR.
CONCLUSION: Considering that HR and MoM-THA used the same MoM bearing design, increased cobalt levels may be related to trunnion wear or corrosion. Elevated cobalt levels should raise concern for corrosion related failure in MoM-THA.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
2024 ACC Expert Consensus Decision Pathway for Treatment of Heart Failure With Reduced Ejection Fraction: A Report of the American College of Cardiology Solution Set Oversight Committee.Journal of the American College of Cardiology 2024 March 3
The Effect of Albumin Administration in Critically Ill Patients: A Retrospective Single-Center Analysis.Critical Care Medicine 2024 Februrary 8
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app