Multicenter study of optical low-coherence interferometry and partial-coherence interferometry optical biometers with patients from the United States and China.
Journal of Cataract and Refractive Surgery 2016 January
PURPOSE: To evaluate the agreement between the measurements provided by a new optical biometer, the Aladdin, based on optical low-coherence interferometry (OLCI), and those provided by the most commonly used optical biometer (IOLMaster 500), based on partial-coherence interferometry (PCI).
SETTING: Multicenter clinical trial.
DESIGN: Prospective evaluation of diagnostic test.
METHODS: In this study, 2 samples of adult patients were enrolled, 1 in the United States and the other in China. The U.S. group included a sample of consecutive patients scheduled for cataract surgery. The China group included a sample of healthy subjects with no cataracts. In both cases, only 1 eye of each patient was analyzed. Axial length (AL), corneal power (in diopters [D]) (K), anterior chamber depth (ACD) (corneal epithelium to lens), and corneal astigmatism were measured. All values were analyzed using a paired t test, the Pearson product-moment correlation coefficient (r), and Bland-Altman plots.
RESULTS: In the U.S. and China groups, the OLCI mean AL values did not show a statistically significant difference from PCI values and showed excellent agreement and correlation. On the contrary, OLCI measured a lower mean K (-0.14 D) and a deeper ACD measurements (U.S. +0.16 mm and China +0.05 mm). These differences were statistically significant (P < .0001). Vector analysis did not show a statistically significant difference in astigmatism measurements.
CONCLUSIONS: Agreement between OLCI and PCI was good. However, the small but statistically significant differences in K and ACD measurements make constant optimization necessary when calculating the intraocular lens power using theoretical formulas.
FINANCIAL DISCLOSURE: Dr. Hoffer licenses the registered trademark name Hoffer to Carl Zeiss-Meditec (PCI), Haag-Streit (Lenstar), Movu (Argos), Oculus (Pentacam, AXL), Nidek (AL-Scan), Tomey (OA-2000), Topcon EU Visia Imaging (Aladdin), Ziemer (Galilei G6), and all A-scan biometer manufacturers. Dr. Shammas licenses his formulas to Carl Zeiss-Meditec (PCI), Haag-Streit (Lenstar), Nidek (AL-Scan), and Topcon EU (Visia Imaging) (Aladdin). None of the other authors has a financial or proprietary interest in any material or method mentioned.
SETTING: Multicenter clinical trial.
DESIGN: Prospective evaluation of diagnostic test.
METHODS: In this study, 2 samples of adult patients were enrolled, 1 in the United States and the other in China. The U.S. group included a sample of consecutive patients scheduled for cataract surgery. The China group included a sample of healthy subjects with no cataracts. In both cases, only 1 eye of each patient was analyzed. Axial length (AL), corneal power (in diopters [D]) (K), anterior chamber depth (ACD) (corneal epithelium to lens), and corneal astigmatism were measured. All values were analyzed using a paired t test, the Pearson product-moment correlation coefficient (r), and Bland-Altman plots.
RESULTS: In the U.S. and China groups, the OLCI mean AL values did not show a statistically significant difference from PCI values and showed excellent agreement and correlation. On the contrary, OLCI measured a lower mean K (-0.14 D) and a deeper ACD measurements (U.S. +0.16 mm and China +0.05 mm). These differences were statistically significant (P < .0001). Vector analysis did not show a statistically significant difference in astigmatism measurements.
CONCLUSIONS: Agreement between OLCI and PCI was good. However, the small but statistically significant differences in K and ACD measurements make constant optimization necessary when calculating the intraocular lens power using theoretical formulas.
FINANCIAL DISCLOSURE: Dr. Hoffer licenses the registered trademark name Hoffer to Carl Zeiss-Meditec (PCI), Haag-Streit (Lenstar), Movu (Argos), Oculus (Pentacam, AXL), Nidek (AL-Scan), Tomey (OA-2000), Topcon EU Visia Imaging (Aladdin), Ziemer (Galilei G6), and all A-scan biometer manufacturers. Dr. Shammas licenses his formulas to Carl Zeiss-Meditec (PCI), Haag-Streit (Lenstar), Nidek (AL-Scan), and Topcon EU (Visia Imaging) (Aladdin). None of the other authors has a financial or proprietary interest in any material or method mentioned.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app