Delayed AICD therapy and cardiac arrest resulting from undersensing of ventricular fibrillation in a subject with hypertrophic cardiomyopathy-A case report.
Indian Pacing and Electrophysiology Journal 2015 March
Defibrillation testing is no longer routinely performed after automatic implantable cardioverter-defibrillator (AICD) implantation. However, certain subjects undergoing AICD implantation may be at higher risk of undersensing of ventricular arrhythmias resulting in potentially fatal outcomes. We present the case of a 30-year-old woman with hypertrophic cardiomyopathy (HCM; 'asymmetric septal hypertophy' morphologic variant) and prophylactic AICD who experienced an out of hospital cardiac arrest. AICD interrogation revealed undersensing as a result of intermittent high amplitude electrograms during an episode of ventricular fibrillation (VF). The subject underwent replacement and repositioning of the AICD lead along with pulse generator replacement (that utilized a different VF sensing algorithm) with appropriate sensing of VF and successful defibrillation testing. The presence of intermittent high amplitude electrograms during episodes of VF in AICDs using the AGC function should be recognized as a situation that may necessitate interventions to prevent undersensing and consequent delay in therapy.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app