Delivery of Epinephrine in the Vapor Phase for the Treatment of Croup.

OBJECTIVES: The Vapotherm system delivers high humidity to the airway of patients by using semipermeable tubules where heated liquid water is in contact with air. The humidified air is conducted to the patient via a heated tube. Preliminary clinical observations in infants with croup suggested that epinephrine added to the water supplying the humidity was delivered successfully in the vapor phase. The purpose of this study was to evaluate the efficiency of the delivery of epinephrine in the vapor phase and to develop the feasibility criteria for a clinical pilot study.

DESIGN: Thirty milligrams of epinephrine in a 1-L bag of sterile water was used as the humidification source for a Vapotherm 2000i. The output of the heated circuit was condensed and collected into a small Erlenmeyer flask via a metal coil while the whole collection system was submerged in an ice slurry to maintain the outflow temperature from the flask between 0°C and 2°C. The in vitro system was tested at 40°C with flows of 5, 10, and 15 L/min and L-epinephrine concentrations of 15, 30, and 60 mg/L. Each test was duplicated at each of the six conditions.

SETTING: Academic children's hospital research laboratory.

PATIENTS: None.

INTERVENTIONS: None.

MEASUREMENTS AND MAIN RESULTS: The system recovered more than 90% of the water vapor from the fully saturated air at 40°C. The epinephrine concentration recovery quantified by ultraviolet-visible spectrophotometry was 23.9% (27.5-20.4%) (mean and range) of the initial concentration. At flows of 5, 10, and 15 L/min, the delivery of epinephrine would be 1.8, 3.6, and 4.2 μg/min, respectively, which is in the therapeutic range used for parenteral infusion in young children.

CONCLUSIONS: The Vapotherm system can be used to deliver epinephrine in pharmacological doses to the respiratory system as a vapor and thus as an alternative to droplets by conventional nebulization.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app