A portable respiratory rate estimation system with a passive single-lead electrocardiogram acquisition module.
Technology and Health Care : Official Journal of the European Society for Engineering and Medicine 2016 July 28
BACKGROUND: Among vital signs of acutely ill hospital patients, respiratory rate (RR) is a highly accurate predictor of health deterioration.
OBJECTIVE: This study proposes a system that consists of a passive and non-invasive single-lead electrocardiogram (ECG) acquisition module and an ECG-derived respiratory (EDR) algorithm in the working prototype of a mobile application.
METHOD: Before estimating RR that produces the EDR rate, ECG signals were evaluated based on the signal quality index (SQI). The SQI algorithm was validated quantitatively using the PhysioNet/Computing in Cardiology Challenge 2011 training data set. The RR extraction algorithm was validated by adopting 40 MIT PhysioNet Multiparameter Intelligent Monitoring in Intensive Care II data set.
RESULTS: The estimated RR showed a mean absolute error (MAE) of 1.4 compared with the ``gold standard'' RR. The proposed system was used to record 20 ECGs of healthy subjects and obtained the estimated RR with MAE of 0.7 bpm.
CONCLUSION: Results indicate that the proposed hardware and algorithm could replace the manual counting method, uncomfortable nasal airflow sensor, chest band, and impedance pneumotachography often used in hospitals. The system also takes advantage of the prevalence of smartphone usage and increase the monitoring frequency of the current ECG of patients with critical illnesses.
OBJECTIVE: This study proposes a system that consists of a passive and non-invasive single-lead electrocardiogram (ECG) acquisition module and an ECG-derived respiratory (EDR) algorithm in the working prototype of a mobile application.
METHOD: Before estimating RR that produces the EDR rate, ECG signals were evaluated based on the signal quality index (SQI). The SQI algorithm was validated quantitatively using the PhysioNet/Computing in Cardiology Challenge 2011 training data set. The RR extraction algorithm was validated by adopting 40 MIT PhysioNet Multiparameter Intelligent Monitoring in Intensive Care II data set.
RESULTS: The estimated RR showed a mean absolute error (MAE) of 1.4 compared with the ``gold standard'' RR. The proposed system was used to record 20 ECGs of healthy subjects and obtained the estimated RR with MAE of 0.7 bpm.
CONCLUSION: Results indicate that the proposed hardware and algorithm could replace the manual counting method, uncomfortable nasal airflow sensor, chest band, and impedance pneumotachography often used in hospitals. The system also takes advantage of the prevalence of smartphone usage and increase the monitoring frequency of the current ECG of patients with critical illnesses.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app