Improving outcomes: case-matched comparison of novel second-generation versus first-generation self-expandable transcatheter heart valves.

OBJECTIVES: The published literature has extensively documented clinical benefit derived from transcatheter aortic valve implantation (TAVI) in high-risk patients using self-expanding current-generation transfemoral (TF) transcatheter heart valves (THVs). However, it has also demonstrated apparent shortcomings such as paravalvular leakage (PVL) or need for permanent pacemaker (PM) implantation. We here present a case-matched analysis of acute 30-day outcomes using a novel nitinol-based THV (Symetis Acurate Neo TF™), which may overcome some limitations of currently used devices.

METHODS: From 2012 to 2015, 69 consecutive patients (study group, 65.2% female, 81.4 ± 6.1 years, logEuroSCORE I 19.9 ± 14.2%) received TF-TAVI using the novel Symetis Acurate Neo TF™ THV. A control group of patients after TF-TAVI with the CoreValve™ THV was retrieved from our database (control group) and matched to the study group utilizing 16 parameters. Data were retrospectively analysed according to updated Valve Academic Research Consortium (VARC-2) definitions. The 30-day follow-up was completed in all cases.

RESULTS: Apart from gender (65.2 vs 44.9% females, P = 0.023), matching was successful with parameters showing no significant differences. The device success rate was 95.6% (66/69) and 89.9% (62/69) in the study and control groups, respectively (P = 0.20). The all-cause 30-day mortality rate was 5.8% (4/69) vs 10.14% (7/69) (P = 0.36), and disabling stroke was observed in 2.9% (2/69) vs 5.8% (4/69) (P = 0.41), respectively. Resultant transvalvular maximum/mean gradient and effective orifice area (EOA) were 13.8 ± 5.5 vs 18.1 ± 8.1 mmHg (P = 0.001)/7.0 ± 2.8 vs 8.8 ± 4.0 mmHg (P = 0.006) and 1.9 ± 0.3 vs 1.8 ± 0.2 cm(2) (P = 0.015), respectively. PVL ≥grade II was observed in 2.9% (2/69) and 15.94% (11/69) (P = 0.013) of patients and the rate of PM implantation was 8.7% (6/69) vs 44.9% (31/69) (P < 0.001), respectively.

DISCUSSION: TF-TAVI was feasible and safe using this new type of nitinol-based THV. Superiority to the current generation of self-expanding THVs was achieved regarding post-interventional pressure gradients and EOA, severity of residual PVL and rate of PM implantation. Results set a promising quality standard for TF-TAVI with a self-expanding THV, but will have to be confirmed in a larger patient cohorts for further clinical evaluation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app