Manual Ventilation and Sustained Lung Inflation in an Experimental Model: Influence of Equipment Type and Operator's Training.

AIM: To compare the influence of devices for manual ventilation and individual experience on the applied respiratory mechanics and sustained lung inflation.

METHODS: A total of 114 instructors and non-instructors from the Neonatal Resuscitation Program of the Brazilian Society of Pediatrics participated in this study. Participants ventilated an intubated manikin. To evaluate respiratory mechanics and sustained lung inflation parameters, a direct comparison was made between the self-inflating bag and the T-shaped resuscitator (T-piece), followed by an analysis of the effectiveness of the equipment according to the participants' education and training.

RESULTS: A difference between equipment types was observed for the tidal volume, with a median (interquartile range) of 28.5 mL (12.6) for the self-inflating bag and 20.1 mL (8.4) for the T-piece in the instructor group and 31.6 mL (14) for the self-inflating bag and 22.3 mL (8.8) for the T-piece in the non-instructor group. Higher inspiratory time values were observed with the T-piece in both groups of professionals, with no significant difference between them. The operator's ability to maintain the target pressure over the 10 seconds of sustained lung inflation was evaluated using the area under the pressure-time curve and was 1.7-fold higher with the use of the T-piece. Inspiratory pressure and mean airway pressure applied during sustained lung inflation were greater with the self-inflating bag, as evaluated between the beginning and the end of the procedure.

CONCLUSION: The T-piece resulted in lower tidal volume and higher inspiratory time values, irrespective of the operator's experience, and increased the ease of performing the sustained lung inflation maneuver, as demonstrated by the maintenance of target pressure for the desired period and a higher mean airway pressure than that obtained using the self-inflating bag.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app