Comparison of Central Corneal Thickness Between Fourier-Domain OCT, Very High-Frequency Digital Ultrasound, and Scheimpflug Imaging Systems.
Journal of Refractive Surgery 2016 Februrary
PURPOSE: To compare corneal thickness measurements between three imaging systems.
METHODS: In this retrospective study of 81 virgin and 58 post-laser refractive surgery corneas, central and minimum corneal thickness were measured using optical coherence tomography (OCT), very high-frequency digital ultrasound (VHF digital ultrasound), and a Scheimpflug imaging system. Agreement between methods was analyzed using mean differences (bias) (OCT - VHF digital ultrasound, OCT - Scheimpflug, VHF digital ultrasound - Scheimpflug) and Bland-Altman analysis with 95% limits of agreement (LoA).
RESULTS: Virgin cornea mean central corneal thickness was 508.3 ± 33.2 µm (range: 434 to 588 µm) for OCT, 512.7 ± 32.2 µm (range: 440 to 587 µm) for VHF digital ultrasound, and 530.2 ± 32.6 µm (range: 463 to 612 µm) for Scheimpflug imaging. OCT and VHF digital ultrasound showed the closest agreement with a bias of -4.37 µm, 95% LoA ±12.6 µm. Least agreement was between OCT and Scheimpflug imaging with a bias of -21.9 µm, 95% LoA ±20.7 µm. Bias between VHF digital ultrasound and Scheimpflug imaging was -17.5 µm, 95% LoA ±19.0 µm. In post-laser refractive surgery corneas, mean central corneal thickness was 417.9 ± 47.1 µm (range: 342 to 557 µm) for OCT, 426.3 ± 47.1 µm (range: 363 to 563 µm) for VHF digital ultrasound, and 437.0 ± 48.5 µm (range: 359 to 571 µm) for Scheimpflug imaging. Closest agreement was between OCT and VHF digital ultrasound with a bias of -8.45 µm, 95% LoA ±13.2 µm. Least agreement was between OCT and Scheimpflug imaging with a bias of -19.2 µm, 95% LoA ±19.2 µm. Bias between VHF digital ultrasound and Scheimpflug imaging was -10.7 µm, 95% LoA ±20.0 µm. No relationship was observed between difference in central corneal thickness measurements and mean central corneal thickness. Results were similar for minimum corneal thickness.
CONCLUSIONS: Central and minimum corneal thickness was measured thinnest by OCT and thickest by Scheimpflug imaging in both groups. A clinically significant bias existed between Scheimpflug imaging and the other two modalities.
METHODS: In this retrospective study of 81 virgin and 58 post-laser refractive surgery corneas, central and minimum corneal thickness were measured using optical coherence tomography (OCT), very high-frequency digital ultrasound (VHF digital ultrasound), and a Scheimpflug imaging system. Agreement between methods was analyzed using mean differences (bias) (OCT - VHF digital ultrasound, OCT - Scheimpflug, VHF digital ultrasound - Scheimpflug) and Bland-Altman analysis with 95% limits of agreement (LoA).
RESULTS: Virgin cornea mean central corneal thickness was 508.3 ± 33.2 µm (range: 434 to 588 µm) for OCT, 512.7 ± 32.2 µm (range: 440 to 587 µm) for VHF digital ultrasound, and 530.2 ± 32.6 µm (range: 463 to 612 µm) for Scheimpflug imaging. OCT and VHF digital ultrasound showed the closest agreement with a bias of -4.37 µm, 95% LoA ±12.6 µm. Least agreement was between OCT and Scheimpflug imaging with a bias of -21.9 µm, 95% LoA ±20.7 µm. Bias between VHF digital ultrasound and Scheimpflug imaging was -17.5 µm, 95% LoA ±19.0 µm. In post-laser refractive surgery corneas, mean central corneal thickness was 417.9 ± 47.1 µm (range: 342 to 557 µm) for OCT, 426.3 ± 47.1 µm (range: 363 to 563 µm) for VHF digital ultrasound, and 437.0 ± 48.5 µm (range: 359 to 571 µm) for Scheimpflug imaging. Closest agreement was between OCT and VHF digital ultrasound with a bias of -8.45 µm, 95% LoA ±13.2 µm. Least agreement was between OCT and Scheimpflug imaging with a bias of -19.2 µm, 95% LoA ±19.2 µm. Bias between VHF digital ultrasound and Scheimpflug imaging was -10.7 µm, 95% LoA ±20.0 µm. No relationship was observed between difference in central corneal thickness measurements and mean central corneal thickness. Results were similar for minimum corneal thickness.
CONCLUSIONS: Central and minimum corneal thickness was measured thinnest by OCT and thickest by Scheimpflug imaging in both groups. A clinically significant bias existed between Scheimpflug imaging and the other two modalities.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app