Characterization and correction of pressure drift in the ManoScan(™) high-resolution manometry system: In vitro and in vivo.

BACKGROUND: A substantial pressure drift in high-resolution manometry (HRM) has been reported; however, fundamental questions remain regarding the origin and management of this drift. The aim of this study was to provide critical in-depth analyses of ManoScan(™) HRM drift in vitro and in vivo.

METHODS: A total of sixteen 15-min studies and twelve 5-h studies were performed in a water bath at 37 °C at 4.0 cm depth (2.9 mmHg) with ESO and ESO Z catheters. Six 5-h in vitro studies were performed similarly at a depth of 9.0 cm (6.6 mmHg). Eight 15-min studies and nine 8-h in vivo studies were performed with healthy participants. Two correction methods - thermal compensation (TC) and interpolated thermal compensation (ITC) - were tested.

KEY RESULTS: Overall pressure drift varied both between studies (p < 0.01) and within sensors (p < 0.01). Drift resulted from thermal shock, an initial pressure change at intubation, and baseline drift, a linear drift over time (R(2)  > 0.96). Contrary to previous reports, there was no correlation between drift and average (r = -0.02) or maximum pressure exposure (r = -0.05). Following data correction, ITC had the lowest median error but persisted with a maximum error of 2.5 mmHg (IQR = 3.0).

CONCLUSIONS & INFERENCES: The substantial drift in the ManoScan(™) HRM system is highly variable and not corrected via the standard operating instructions. ITC has superior performance but requires communication with the manufacturer to enable this option. This has a substantial impact on clinical diagnosis, utility of existing normative data, and future research of HRM.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app