Evaluation Study
Journal Article
Add like
Add dislike
Add to saved papers

Repeatability of posterior and total corneal curvature measurements with a dual Scheimpflug-Placido tomographer.

PURPOSE: To evaluate the repeatability of the Galilei G4 dual Scheimpflug analyzer in measuring simulated keratometric, total, and posterior corneal curvature in normal and post-refractive surgery eyes.

SETTING: Cullen Eye Institute, Baylor College of Medicine, Houston, Texas, USA.

DESIGN: Prospective evaluation of diagnostic technology.

METHODS: A single observer performed 3 consecutive measurements in 1 eye of each subject. The following were evaluated in both eyes and in eyes that had previous myopic excimer-laser surgery: (1) simulated keratometric corneal power and astigmatism, (2) total corneal power and astigmatism, and (3) posterior corneal power and astigmatism. Repeatability was assessed by calculating the within-subject standard deviation (Sw), coefficient of variation (CoV), and intra-class correlation coefficient (ICC).

RESULTS: The study evaluated 41 normal eyes and 36 post-refractive surgery eyes. In normal eyes, the Sw was 0.08 diopters (D), 0.10 D, and 0.03 D for simulated keratometric, total, and posterior corneal power, respectively. The CoV ranged from 0.16% to 0.40%, and the ICC was 0.992 or more (P < .001) for all corneal powers. In post-refractive surgery eyes, the Sw was 0.09 D, 0.09 D, and 0.02 D for simulated keratometric, total, and posterior corneal power, respectively. The CoV ranged from 0.19% to 0.32%, and the ICC was 0.990 or more (P < .001) for all corneal powers. For posterior corneal astigmatism, the ICC was 0.814 and 0.886 for normal and post-refractive surgery eyes, respectively.

CONCLUSIONS: In normal corneas and corneas that had undergone myopic excimer laser ablation, the dual Scheimpflug analyzer showed high intra-observer repeatability for simulated keratometric, total, and posterior corneal power measurements and moderate repeatability for posterior corneal astigmatism.

FINANCIAL DISCLOSURE: Drs. Koch, Weikert, and Wang received research support from Ziemer USA, Inc. No other author has a financial or proprietary interest in any material or method mentioned.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app