Quantitative analysis of brass compensators for commissioning of the Pinnacle planning system for IMRT.
Journal of Applied Clinical Medical Physics 2015 November 9
Brass compensators for beam modulation present an alternative method for IMRT treatment planning to traditional dynamic leaf modulation. In this work, we present a detailed method to commission the Pinnacle treatment planning system for IMRT using brass compensators. Beam attenuation from various brass thicknesses were measured using an ion chamber, as well as a MapCHECK device, for a representative seven-field IMRT plan. We show that Pinnacle's parameters for compensators can be optimized to match the measured results while still maintaining the original open-beam model. Only Pinnacle's modified scatter factor and brass density value were adjusted. Beam attenuation in several brass slabs were optimized by minimizing a χ(2) function of measured and modeled data for several depths in solid water. Using MapCHECK, Pinnacle's parameters were optimized by minimizing the number of points that fail during a gamma analysis of seven fields in a typical IMRT plan. Our results show that Pinnacle's 10X beam is best modeled using a brass density of 7.8 g/cm(3) and a modified scatter factor of 0.1 cm(-1).
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app