Effects of Rotational Motion in Robotic Needle Insertion.
Journal of Biomedical Physics & Engineering 2015 December
BACKGROUND: Robotic needle insertion in biological tissues has been known as one the most applicable procedures in sampling, robotic injection and different medical therapies and operations.
OBJECTIVE: In this paper, we would like to investigate the effects of angular velocity in soft tissue insertion procedure by considering force-displacement diagram. Non-homogenous camel liver can be exploited as a tissue sample under standard compression test with Zwick/Roell device employing 1-D axial load-cell.
METHODS: Effects of rotational motion were studied by running needle insertion experiments in 5, 50 and 200 mm/min in two types of with or without rotational velocity of 50, 150 and 300 rpm. On further steps with deeper penetrations, friction force of the insertion procedure in needle shaft was acquired by a definite thickness of the tissue.
RESULTS: Designed mechanism of fixture for providing different frequencies of rotational motion is available in this work. Results for comparison of different force graphs were also provided.
CONCLUSION: Derived force-displacement graphs showed a significant difference between two procedures; however, tissue bleeding and disorganized micro-structure would be among unavoidable results.
OBJECTIVE: In this paper, we would like to investigate the effects of angular velocity in soft tissue insertion procedure by considering force-displacement diagram. Non-homogenous camel liver can be exploited as a tissue sample under standard compression test with Zwick/Roell device employing 1-D axial load-cell.
METHODS: Effects of rotational motion were studied by running needle insertion experiments in 5, 50 and 200 mm/min in two types of with or without rotational velocity of 50, 150 and 300 rpm. On further steps with deeper penetrations, friction force of the insertion procedure in needle shaft was acquired by a definite thickness of the tissue.
RESULTS: Designed mechanism of fixture for providing different frequencies of rotational motion is available in this work. Results for comparison of different force graphs were also provided.
CONCLUSION: Derived force-displacement graphs showed a significant difference between two procedures; however, tissue bleeding and disorganized micro-structure would be among unavoidable results.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app