Adjusting ventilator off-cycling in invasively ventilated COPD patients needs comprehensive adjustments.
Minerva Anestesiologica 2016 July
BACKGROUND: Patient-ventilator asynchrony that prolongs weaning and increases morbidity and mortality is common during invasive ventilation of patients with chronic obstructive pulmonary disease (COPD). In this context, the inspiratory cycling criteria (iCC) of the ventilator during assisted pressure support (PS) ventilation is a poorly acknowledged key factor. We investigated the changes of flow and pressure parameters that resulted from varying the iCC in a simulated COPD lung model.
METHODS: A lung simulator was connected to an ICU ventilator through an endotracheal tube. We studied iCC settings from 10% to 70% at different respiratory rates (RR) (15 and 30 bpm) and pressure support (PS) (5 and 15 cmH2O) settings and registered asynchrony-index, double-triggering, expiratory trigger latency (TLEXP), intrinsic PEEP (PEEPi), expiratory pressure time product (PTPEXP) and tidal volume.
RESULTS: At iCC≤20%, asynchrony occurred in 50% of all recordings in high RR/high PS. At a low RR, double triggering occurred at high iCC settings. It appeared at 50% iCC with low PS and at 60% iCC with high PS. TLEXP was positive at iCC 10% to 30% but decreased with increasing iCC (P<0.001). At low RR/high PS settings, PEEPi decreased at iCC≤40% but increased at iCC≥50%. High RR/low PS constantly reduced PTPEXP up to 60% iCC. Changes in iCC strongly influenced the resulting tidal volume.
CONCLUSIONS: Non-adapted ventilator iCC can cause patient-ventilator asynchrony. The success of assisted invasive ventilation and weaning relies on meticulous adjustments.
METHODS: A lung simulator was connected to an ICU ventilator through an endotracheal tube. We studied iCC settings from 10% to 70% at different respiratory rates (RR) (15 and 30 bpm) and pressure support (PS) (5 and 15 cmH2O) settings and registered asynchrony-index, double-triggering, expiratory trigger latency (TLEXP), intrinsic PEEP (PEEPi), expiratory pressure time product (PTPEXP) and tidal volume.
RESULTS: At iCC≤20%, asynchrony occurred in 50% of all recordings in high RR/high PS. At a low RR, double triggering occurred at high iCC settings. It appeared at 50% iCC with low PS and at 60% iCC with high PS. TLEXP was positive at iCC 10% to 30% but decreased with increasing iCC (P<0.001). At low RR/high PS settings, PEEPi decreased at iCC≤40% but increased at iCC≥50%. High RR/low PS constantly reduced PTPEXP up to 60% iCC. Changes in iCC strongly influenced the resulting tidal volume.
CONCLUSIONS: Non-adapted ventilator iCC can cause patient-ventilator asynchrony. The success of assisted invasive ventilation and weaning relies on meticulous adjustments.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app