A posterior anal sling for fecal incontinence: results of a 152-patient prospective multicenter study.

BACKGROUND: The transobturator posterior anal sling (TOPAS) system is a posterior anal sling that is a minimally invasive, self-fixating polypropylene mesh intended to treat fecal incontinence (FI) in women who have failed conservative therapy.

OBJECTIVE: We are reporting 1-year outcome in a prospective, multicenter study under investigational device exemption, evaluating this new treatment modality.

STUDY DESIGN: A total of 152 women were implanted with the TOPAS system at 14 centers in the United States. FI was assessed preoperatively and at the 12-month follow up with a 14-day bowel diary, Cleveland Clinic Incontinence Scores, and FI Quality of Life questionnaires. Treatment success was defined as reduction in number of FI episodes of ≥50% compared to baseline. Missing bowel diary data were considered treatment failures. The Wilcoxon signed rank test was used to compare changes observed at 12 months vs baseline.

RESULTS: Mean age was 59.6 years old (SD 9.7). The mean duration of FI was 110 mo (range 8-712) months. Mean length of the implant procedure was 33.4 (SD 11.6) minutes. Mean EBL was 12.9 (SD 10.5) mL. Average follow-up was 24.9 months. At 12 months, 69.1% of patients met the criteria for treatment success, and 19% of subjects reported complete continence. FI episodes/wk decreased from a median of 9.0 (range 2-40) at baseline to 2.5 (range 0-40) (P < .001). FI days decreased from a median of 5.0 (range 1.5-7) at baseline to 2.0 (range 0-7) (P < .001) over a 7-day period. FI associated with urgency decreased from a median at baseline of 2.0 (range 0-26) to 0 (range 0-14.5) (P < .001). The mean Cleveland Clinic Incontinence Scores decreased from 13.9 at baseline to 9.6 at 12 months (P < .001). FI Quality of Life scores for all 4 domains improved significantly from baseline to 12 months (P < .001). A total of 66 subjects experienced 104 procedure- and/or device-related adverse events (AEs). Most AEs were short in duration and 97% were managed without therapy or with nonsurgical interventions. No treatment-related deaths, erosions, extrusions, or device revisions were reported. The most common AE categories were pelvic pain (n = 47) and infection (n = 26). Those subjects experiencing pelvic pain had a mean pain score (0-10 scale, 0 = no pain) during the 12-month follow-up of 1.2 (SD 2.4).

CONCLUSION: The TOPAS system provides significant improvements in FI symptoms and quality of life with an acceptable AE profile and may therefore be a viable minimally invasive treatment option for FI in women.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app