Does superficial fat affect metabolite concentrations determined by MR spectroscopy with water referencing?

NMR in Biomedicine 2015 November
It has recently been reported in this journal that local fat depots produce a sizable frequency-dependent signal attenuation in magnetic resonance spectroscopy (MRS) of the brain. If of a general nature, this effect would question the use of internal reference signals for quantification of MRS and the quantitative use of MRS as a whole. Here, it was attempted to verify this effect and pinpoint the potential causes by acquiring data with various acquisition settings, including two field strengths, two MR scanners from different vendors, different water suppression sequences, RF coils, localization sequences, echo times, and lipid/metabolite phantoms. With all settings tested, the reported effect could not be reproduced, and it is concluded that water referencing and quantitative MRS per se remain valid tools under common acquisition conditions.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app