Flow dynamics in pediatric rigid bronchoscopes using computer-aided design modeling software.
Laryngoscope 2016 August
OBJECTIVES/HYPOTHESIS: Observed complications during rigid bronchoscopy, including hypercarbia and hypoxemia, prompted us to assess how well rigid bronchoscopes serve as an airway device. We performed computer-aided design flow analysis of pediatric rigid bronchoscopes to gain insight into flow dynamics.
STUDY DESIGN: We made accurate three-dimensional computer models of pediatric rigid bronchoscopes and endotracheal tubes. SOLIDWORKS (Dassault Systemes, Vélizy-Villacoublay, France) flow analysis software was used to analyze fluid dynamics during pressure-controlled and volume-controlled ventilation.
METHODS: Flow analysis was performed on rigid bronchoscopes and similar outer diameter endotracheal tubes comparing resistance, flow, and turbulence during two ventilation modalities and in common surgical scenarios.
RESULTS: Increased turbulent flow was observed in bronchoscopes compared to more laminar flow in endotracheal tubes of similar outer diameter. Flow analysis displayed higher resistances in all pediatric bronchoscope sizes except one (3.0 bronchoscope) compared to similar-sized endotracheal tubes. Loss of adequate ventilation was observed if the bronchoscope was not assembled correctly or if increased peak inspiratory pressures were needed. Anesthesia flow to the patient was reduced by 63% during telescope insertion.
CONCLUSIONS: Flow analysis illustrates increased turbulent flow and increased airflow resistance in all but one size of pediatric bronchoscopes compared to endotracheal tubes. This increased turbulence and resistance, along with the unanticipated gas distal exit pattern, may contribute to the documented hypercarbia and hypoxemia during procedures. These findings may explain why hypoxemia and hypercarbia are commonly observed during rigid bronchoscopy, especially when positive pressure ventilation is needed.
LEVEL OF EVIDENCE: NA Laryngoscope, 126:1940-1945, 2016.
STUDY DESIGN: We made accurate three-dimensional computer models of pediatric rigid bronchoscopes and endotracheal tubes. SOLIDWORKS (Dassault Systemes, Vélizy-Villacoublay, France) flow analysis software was used to analyze fluid dynamics during pressure-controlled and volume-controlled ventilation.
METHODS: Flow analysis was performed on rigid bronchoscopes and similar outer diameter endotracheal tubes comparing resistance, flow, and turbulence during two ventilation modalities and in common surgical scenarios.
RESULTS: Increased turbulent flow was observed in bronchoscopes compared to more laminar flow in endotracheal tubes of similar outer diameter. Flow analysis displayed higher resistances in all pediatric bronchoscope sizes except one (3.0 bronchoscope) compared to similar-sized endotracheal tubes. Loss of adequate ventilation was observed if the bronchoscope was not assembled correctly or if increased peak inspiratory pressures were needed. Anesthesia flow to the patient was reduced by 63% during telescope insertion.
CONCLUSIONS: Flow analysis illustrates increased turbulent flow and increased airflow resistance in all but one size of pediatric bronchoscopes compared to endotracheal tubes. This increased turbulence and resistance, along with the unanticipated gas distal exit pattern, may contribute to the documented hypercarbia and hypoxemia during procedures. These findings may explain why hypoxemia and hypercarbia are commonly observed during rigid bronchoscopy, especially when positive pressure ventilation is needed.
LEVEL OF EVIDENCE: NA Laryngoscope, 126:1940-1945, 2016.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app