We have located links that may give you full text access.
Accuracy of different types of computer-aided design/computer-aided manufacturing surgical guides for dental implant placement.
PURPOSE: To evaluate the clinical outcomes of implants placed using different types of computer-aided design/computer-aided manufacturing (CAD/CAM) surgical guides, including partially guided and totally guided templates, and determine the accuracy of these guides Materials and methods: In total, 111 implants were placed in 24 patients using CAD/CAM surgical guides. After implant insertion, the positions and angulations of the placed implants relative to those of the planned ones were determined using special software that matched pre- and postoperative computed tomography (CT) images, and deviations were calculated and compared between the different guides and templates.
RESULTS: The mean angular deviations were 1.72 ± 1.67 and 2.71 ± 2.58, the mean deviations in position at the neck were 0.27 ± 0.24 and 0.69 ± 0.66 mm, the mean deviations in position at the apex were 0.37 ± 0.35 and 0.94 ± 0.75 mm, and the mean depth deviations were 0.32 ± 0.32 and 0.51 ± 0.48 mm with tooth- and mucosa-supported stereolithographic guides, respectively (P < .05 for all). The mean distance deviations when partially guided (29 implants) and totally guided templates (30 implants) were used were 0.54 ± 0.50 mm and 0.89 ± 0.78 mm, respectively, at the neck and 1.10 ± 0.85 mm and 0.81 ± 0.64 mm, respectively, at the apex, with corresponding mean angular deviations of 2.56 ± 2.23° and 2.90 ± 3.0° (P > .05 for all).
CONCLUSIONS: Tooth-supported surgical guides may be more accurate than mucosa-supported guides, while both partially and totally guided templates can simplify surgery and aid in optimal implant placement.
RESULTS: The mean angular deviations were 1.72 ± 1.67 and 2.71 ± 2.58, the mean deviations in position at the neck were 0.27 ± 0.24 and 0.69 ± 0.66 mm, the mean deviations in position at the apex were 0.37 ± 0.35 and 0.94 ± 0.75 mm, and the mean depth deviations were 0.32 ± 0.32 and 0.51 ± 0.48 mm with tooth- and mucosa-supported stereolithographic guides, respectively (P < .05 for all). The mean distance deviations when partially guided (29 implants) and totally guided templates (30 implants) were used were 0.54 ± 0.50 mm and 0.89 ± 0.78 mm, respectively, at the neck and 1.10 ± 0.85 mm and 0.81 ± 0.64 mm, respectively, at the apex, with corresponding mean angular deviations of 2.56 ± 2.23° and 2.90 ± 3.0° (P > .05 for all).
CONCLUSIONS: Tooth-supported surgical guides may be more accurate than mucosa-supported guides, while both partially and totally guided templates can simplify surgery and aid in optimal implant placement.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app