[Analysis of Pulse Rate Variability and Its Application to Wearable Smart Devices].

In this paper, a reflection type photoelectric pulse wave sensor was designed for short-term pulse rate variability analysis. Photoplethysmography (PPG) signals and ECG signals (obtained with the Dimetek MicroECG recorder Dicare-m1CP) were recorded synchronously from 20 healthy subjects. The analytical results show a significant correlation (correlation coefficient r > 0.99) between the PPG-derived peak-to-peak (PP) intervals and the ECG-derived RR intervals. Besides, there are no significant differences (P > 0.05) between the HRV measured by ECG and the PRV quantified by the PPG whether in time domain, frequency domain, or the Poincaré plot parameters. The experimental results suggest that the PPG-based short-term PRV analysis can be consistent with the ECG-based HRV measurement in wearable smart devices.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app