Effectiveness of electrocardiographic guidance in CVAD tip placement.
British Journal of Nursing : BJN 2015 July 24
BACKGROUND: International standard practice for the correct confirmation of the central venous access device is the chest X-ray. The intracavitary electrocardiogram-based insertion method is radiation-free, and allows real-time placement verification, providing immediate treatment and reduced requirement for post-procedural repositioning.
METHODS: Relevant databases were searched for prospective randomised controlled trials (RCTs) or quasi RCTs that compared the effectiveness of electrocardiogram-guided catheter tip positioning with placement using surface-anatomy-guided insertion plus chest X-ray confirmation. The primary outcome was accurate catheter tip placement. Secondary outcomes included complications, patient satisfaction and costs.
RESULTS: Five studies involving 729 participants were included. Electrocardiogram-guided insertion was more accurate than surface anatomy guided insertion (odds ratio: 8.3; 95% confidence interval (CI) 1.38; 50.07; p=0.02). There was a lack of reporting on complications, patient satisfaction and costs.
CONCLUSION: The evidence suggests that intracavitary electrocardiogram-based positioning is superior to surface-anatomy-guided positioning of central venous access devices, leading to significantly more successful placements. This technique could potentially remove the requirement for post-procedural chest X-ray, especially during peripherally inserted central catheter (PICC) line insertion.
METHODS: Relevant databases were searched for prospective randomised controlled trials (RCTs) or quasi RCTs that compared the effectiveness of electrocardiogram-guided catheter tip positioning with placement using surface-anatomy-guided insertion plus chest X-ray confirmation. The primary outcome was accurate catheter tip placement. Secondary outcomes included complications, patient satisfaction and costs.
RESULTS: Five studies involving 729 participants were included. Electrocardiogram-guided insertion was more accurate than surface anatomy guided insertion (odds ratio: 8.3; 95% confidence interval (CI) 1.38; 50.07; p=0.02). There was a lack of reporting on complications, patient satisfaction and costs.
CONCLUSION: The evidence suggests that intracavitary electrocardiogram-based positioning is superior to surface-anatomy-guided positioning of central venous access devices, leading to significantly more successful placements. This technique could potentially remove the requirement for post-procedural chest X-ray, especially during peripherally inserted central catheter (PICC) line insertion.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app