Breast cancer detection rates using four different types of mammography detectors.

OBJECTIVE: To compare the performance of different types of detectors in breast cancer detection.

METHODS: A mammography image set containing subtle malignant non-calcification lesions, biopsy-proven benign lesions, simulated malignant calcification clusters and normals was acquired using amorphous-selenium (a-Se) detectors. The images were adapted to simulate four types of detectors at the same radiation dose: digital radiography (DR) detectors with a-Se and caesium iodide (CsI) convertors, and computed radiography (CR) detectors with a powder phosphor (PIP) and a needle phosphor (NIP). Seven observers marked suspicious and benign lesions. Analysis was undertaken using jackknife alternative free-response receiver operating characteristics weighted figure of merit (FoM). The cancer detection fraction (CDF) was estimated for a representative image set from screening.

RESULTS: No significant differences in the FoMs between the DR detectors were measured. For calcification clusters and non-calcification lesions, both CR detectors' FoMs were significantly lower than for DR detectors. The calcification cluster's FoM for CR NIP was significantly better than for CR PIP. The estimated CDFs with CR PIP and CR NIP detectors were up to 15% and 22% lower, respectively, than for DR detectors.

CONCLUSION: Cancer detection is affected by detector type, and the use of CR in mammography should be reconsidered.

KEY POINTS: The type of mammography detector can affect the cancer detection rates. CR detectors performed worse than DR detectors in mammography. Needle phosphor CR performed better than powder phosphor CR. Calcification clusters detection is more sensitive to detector type than other cancers.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

Managing Alcohol Withdrawal Syndrome.Annals of Emergency Medicine 2024 March 26

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app