Outcomes for Myopic LASIK With the MEL 90 excimer laser.

PURPOSE: To evaluate the visual outcomes of myopic LASIK performed with the MEL 90 excimer laser (Carl Zeiss Meditec AG, Jena, Germany) using the Triple-A profile with a 500-Hz pulse rate.

METHODS: Retrospective analysis of the first 286 myopic LASIK procedures (147 patients) by two experienced surgeons in which the VisuMax femtosecond laser and MEL 90 excimer laser (Carl Zeiss Meditec) were used following a standardized surgical technique. Inclusion criteria were preoperative spherical equivalent refraction (SEQ) up to -10.38 diopters (D), cylinder up to 5.00 D, and corrected distance visual acuity (CDVA) of 20/25 or better. No nomogram adjustments were made. Patients were observed for 3 months. Flap thickness was between 80 and 110 µm and optical zone was between 6 and 7 mm. Standard outcomes analysis was performed.

RESULTS: Preoperatively, mean SEQ was -3.83 ± 1.83 D (range: -0.13 to -10.38 D) and mean cylinder was -0.94 ± 0.86 D (range: 0.00 to -5.00 D). Mean age was 36.4 years (range: 18.2 to 74.1 years) with 50% female patients. Of this population, 138 eyes were treated by one surgeon and 148 eyes by another. The mean predictability of SEQ was -0.13 ± 0.34 D (range: -1.00 to +1.00 D). Postoperative SEQ was ± 0.50 D in 88% and ± 1.00 D in 100% of eyes. Preoperative CDVA was 20/20 or better in 97% of eyes. Postoperative uncorrected distance visual acuity was 20/20 or better in 92% and 20/25 or better in 99% of eyes. One line of CDVA was lost in 6% of eyes and no eyes lost two or more lines. There was statistically significant improvement in mesopic contrast sensitivity (CSV-1000) at 3 (P = .021), 6, 12, and 18 (all P ≤.001) cycles per degree.

CONCLUSIONS: The MEL 90 excimer laser using the Triple-A ablation profile with a 500-Hz pulse rate was found to achieve a small but real increase in contrast sensitivity and high efficacy for myopia up to -10.00 D and cylinder up to 5.00 D without the need for a nomogram adjustment.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app