Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography.

AIM: To compare the anterior segment measurements obtained by rotating Scheimpflug camera (Pentacam) and Scanning-slit topography (Orbscan IIz) in keratoconic eyes.

METHODS: A total of 121 patients, 71 males (58.7%) and 50 females (41.3%) (214 eyes) with the diagnosis of keratoconus (KC) were enrolled in this study. Following diagnosis of KC by slit-lamp biomicroscopic examination, central corneal thickness (CCT), thinnest corneal thickness (TCT), anterior chamber depth (ACD), and pupil diameter (PD) were measured by a single examiner using successive instrumentation by Pentacam and Orbscan.

RESULTS: There was no significant difference between the two instruments for the measurement of CCT and TCT. In contrast, scanning-slit topography measured ACD (3.46±0.40 mm vs. 3.38±0.33 mm, P=0.019) and PD (4.97±1.26 mm vs 4.08±1.19 mm, P<0.001) significantly larger than rotating Scheimpflug camera. The two devices made similar measurements for CCT (95% CI: -2.94 to 5.06, P=0.602). However, the mean difference for TCT was -6.28 (95% CI: -10.51 to -2.06, P=0.004) showing a thinner measurement by Orbscan than by Pentacam. In terms of the ACD, the mean difference was 0.08 mm (95% CI: 0.04 to 0.12, P<0.001) with Orbscan giving a slightly larger value than Pentacam. Similarly, Orbscan measurement for PD was longer than Pentacam (95% CI: 0.68 to 1.08, P<0.001).

CONCLUSION: A good agreement was found between Pentacam and Orbscan concerning CCT measurement while comparing scanning-slit topography and rotating Scheimpflug camera there was an underestimation for TCT and overestimation for ACD and PD.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app