Comparison of central and peripheral corneal thickness measurements with scanning-slit, Scheimpflug and Fourier-domain ocular coherence tomography.

PURPOSE: To compare central, regional and relational corneal thickness values obtained with multiple technologies in normal patients and to determine their equivalence and interchangeability.

METHODS: Retrospective analysis of 100 eyes from 50 patients evaluated by ultrasound pachymetry (Pachette II), scanning-slit (Orbscan II), Scheimpflug (Pentacam HR) and spectral-domain ocular coherence tomography (OCT) (RTVue-100) obtained as average values (OCT-A) and point measurements (OCT-P). Measurements included central corneal thickness (CCT) for all technologies and thinnest corneal thickness for scanning-slit, Scheimpflug and OCT. Peripheral thickness measurements were obtained at the 3 mm radius in the superior (S), nasal (N), inferior (I) and temporal (T) regions.

RESULTS: CCT values were: 563.9±36.1μ ultrasound, 570.9±36.1μ scanning-slit, 552.8±33.8μ Scheimpflug, 550.5±32.7μ (OCT-A), 549.4±32.7μ (OCT-P). Ultrasound and scanning-slit were significantly different from each other (p<0.0001), and both were significantly different from all other devices (p<0.0001), while Scheimpflug was similar to OCT-A and OCT-P (p=0.4). Differences between CCT and thinnest corneal thickness were significantly different from all technologies except scanning-slit and OCT-A. For peripheral values, almost all locations' measurements were significantly different from one another (p<0.0001). Superior-inferior values and ratios were also significantly different from one another for almost all devices with no consistent patterns detectible.

CONCLUSIONS: There are significant clinically relevant differences between regional and relational thickness measurements obtained with ultrasound, scanning-slit, Scheimpflug and OCT devices. Screening metrics devised for one system do not appear directly applicable to other measurement systems.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app