Case Reports
Journal Article
Technical Report
Add like
Add dislike
Add to saved papers

Microcatheter Shaping for Intracranial Aneurysm Coiling Using the 3-Dimensional Printing Rapid Prototyping Technology: Preliminary Result in the First 10 Consecutive Cases.

OBJECTIVE: An optimal microcatheter is necessary for successful coiling of an intracranial aneurysm. The optimal shape may be predetermined before the endovascular surgery via the use of a 3-dimensional (3D) printing rapid prototyping technology. We report a preliminary series of intracranial aneurysms treated with a microcatheter shape determined by the patient's anatomy and configuration of the aneurysm, which was fabricated with a 3D printer aneurysm model.

METHODS: A solid aneurysm model was fabricated with a 3D printer based on the data acquired from the 3D rotational angiogram. A hollow aneurysm model with an identical vessel and aneurysm lumen to the actual anatomy was constructed with use of the solid model as a mold. With use of the solid model, a microcatheter shaping mandrel was formed to identically line the 3D curvature of the parent vessel and the long axis of the aneurysm. With use of the mandrel, a test microcatheter was shaped and validated for the accuracy with the hollow model. All the planning processes were undertaken at least 1 day before treatment. The preshaped mandrel was then applied in the endovascular procedure. Ten consecutive intracranial aneurysms were coiled with the pre-planned shape of the microcatheter and evaluated for the clinical and anatomical outcomes and microcatheter accuracy and stability.

RESULTS: All of pre-planned microcatheters matched the vessel and aneurysm anatomy. Seven required no microguidewire assistance in catheterizing the aneurysm whereas 3 required guiding of a microguidewire. All of the microcatheters accurately aligned the long axis of the aneurysm. The pre-planned microcatheter shapes demonstrated stability in all except in 1 large aneurysm case.

CONCLUSION: When a 3D printing rapid type prototyping technology is used, a patient-specific and optimal microcatheter shape may be determined preoperatively.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app