Biomechanical study of dynamic changes in L4-L5 foramen surface area in flexion and extension after implantation of four interspinous process devices.

BACKGROUND: Lumbar spinal stenosis is a major public health issue. Interspinous devices implanted using minimally invasive techniques may constitute an alternative to the reference standard of bony decompression with or without intervertebral fusion. However, their indications remain unclear, due to a paucity of clinical and biomechanical data. Our objective was to evaluate the effects of four interspinous process devices implanted at L4-L5 on the intervertebral foramen surface areas at the treated and adjacent levels, in flexion and in extension.

MATERIALS AND METHOD: Six fresh frozen human cadaver lumbar spines (L2-sacrum) were tested on a dedicated spinal loading frame, in flexion and extension, from 0 to 10 N·m, after preparation and marking of the L3-L4, L4-L5, and L5-S1 foramina. Stereoscopic 3D images were acquired at baseline then after implantation at L4-L5 of each of the four devices (Inspace(®), Synthes; X-Stop(®), Medtronic; Wallis(®), Zimmer; and Diam(®), Medtronic). The surface areas of the three foramina of interest were computed.

RESULTS: All four devices significantly opened the L4-L5 foramen in extension. The effects in flexion separated the devices into two categories. With the two devices characterized by fixation in the spinous processes (Wallis(®) and Diam(®)), the L4-L5 foramen opened only in extension; whereas with the other two devices (X-Stop(®) and Inspace(®)), the L4-L5 foramen opened not only in extension, but also in flexion and in the neutral position. None of the devices implanted at L4-L5 modified the size of the L3-L4 foramen. X-Stop(®) and Diam(®) closed the L5-S1 foramen in extension, whereas the other two devices had no effect at this level.

CONCLUSION: Our results demonstrate that interspinous process devices modify the surface area of the interspinous foramina in vitro. Clinical studies are needed to clarify patient selection criteria for interspinous process device implantation.

LEVEL OF EVIDENCE: Level IV. Investigating an orthopaedic device.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app