Macroscopic barotrauma caused by stiff and soft-tipped airway exchange catheters: an in vitro case series.

BACKGROUND: Many airway management guidelines include the use of airway exchange catheters (AECs). There are reports, however, of harm from their use, from both malpositioning and in particular from the administration of oxygen via an AEC leading to barotrauma.

METHODS: We used an in vitro pig lung model to investigate the safety of administering oxygen at 4 different flow rates from a high-pressure source via 2 different AECs: a standard catheter and a soft-tipped catheter. Experiments were performed with the catheters positioned either above the carina or below it at the first point of resistance to advancement (hold-up). The experiments were then repeated to produce a series of 32 cases.

RESULTS: With an AEC positioned above the carina, we did not observe macroscopic lung damage after the administration of oxygen. The administration of oxygen through an AEC positioned below the carina resulted in macroscopic barotrauma regardless of the rate of oxygen delivery. Increasing speed of oxygen flow led to faster and more extensive damage. Use of an "injector" at 2.5 or 4 bar led to instantaneous macroscopic lung damage and advancement of the AEC through the lung tissue. Our observations were the same when both types of AECs were used.

CONCLUSIONS: Our results are consistent with reports of harm during the use of AECs and demonstrate the risk of administering oxygen through these devices when they are positioned below the carina. An indicator, ideally made on an AEC at the time of manufacture and designed to lie at the same level as the teeth, may be useful in preventing the insertion of that AEC beyond the level of the carina and improve the safety of using such devices.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app