Comparative Study
Journal Article
Research Support, Non-U.S. Gov't
Add like
Add dislike
Add to saved papers

An in vitro comparison of microbial ingress into 8 different needleless IV access devices.

There are conflicting reports of the effect needleless intravenous access devices have on rates of catheter-related bloodstream infection. The aim of this study was to identify any differences between the rates of microbial ingress into 8 different devices following contamination. Each type of device was subjected to a 7-day clinical simulation that involved repeated microbial contamination of the injection site and decontamination followed by saline flushes. Significant differences in the number of microorganisms associated with each device were detected in the saline eluates. Three positive-displacement mechanical valves were associated with the ingress of significantly fewer microorganisms compared with other devices.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app