Prevention of airway fires: testing the safety of endotracheal tubes and surgical devices in a mechanical model.
American Journal of Otolaryngology 2015 January
OBJECTIVES: This study was designed to assess the ability of carbon dioxide (CO2) lasers and radiofrequency ablation devices (Coblator) (ArthoCare Corporation, Sunnyvale, CA) to ignite either a non-reinforced (polyvinylchloride) endotracheal tube (ETT) or an aluminum and fluoroplastic wrapped silicon ("laser safe") ETT at varying titrations of oxygen in a mechanical model of airway surgery.
METHODS: Non-reinforced and laser safe ETTs were suspended in a mechanical model imitating endoscopic airway surgery. A CO2 laser set at 5-30 watts was fired at the ETT at oxygen concentrations ranging from 21% to 88%. The process was repeated using a radiofrequency ablation (RFA) device. All trials were repeated to ensure accuracy.
RESULTS: The CO2 laser ignited a fire when contacting a non-reinforced ETT in under 2 seconds at oxygen concentrations as low as 44%. The CO2 laser could not ignite a laser safe ETT under any conditions, unless it struck the non-reinforced distal tip of the ETT. With the RFA, a fire could not be ignited with either reinforced or non-reinforced ETTs.
CONCLUSIONS: RFA presents no risk of ignition in simulated airway surgery. CO2 lasers should be utilized with a reinforced ETT or no ETT, as fires can easily ignite when lasers strike a non-reinforced ETT. Decreasing the fraction of inspired oxygen reduces the risk of fire.
METHODS: Non-reinforced and laser safe ETTs were suspended in a mechanical model imitating endoscopic airway surgery. A CO2 laser set at 5-30 watts was fired at the ETT at oxygen concentrations ranging from 21% to 88%. The process was repeated using a radiofrequency ablation (RFA) device. All trials were repeated to ensure accuracy.
RESULTS: The CO2 laser ignited a fire when contacting a non-reinforced ETT in under 2 seconds at oxygen concentrations as low as 44%. The CO2 laser could not ignite a laser safe ETT under any conditions, unless it struck the non-reinforced distal tip of the ETT. With the RFA, a fire could not be ignited with either reinforced or non-reinforced ETTs.
CONCLUSIONS: RFA presents no risk of ignition in simulated airway surgery. CO2 lasers should be utilized with a reinforced ETT or no ETT, as fires can easily ignite when lasers strike a non-reinforced ETT. Decreasing the fraction of inspired oxygen reduces the risk of fire.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app