ENGLISH ABSTRACT
JOURNAL ARTICLE
RANDOMIZED CONTROLLED TRIAL
Add like
Add dislike
Add to saved papers

[Study of low radiation exposure dose and low contrast medium dose in coronary CT angiography with High-pitch spiral acquisition mode of dual source CT].

OBJECTIVE: To evaluate the feasibility of low radiation exposure and low contrast medium volume for coronary CT angiography with High- pitch spiral acquisition mode of dual source CT.

METHODS: 135 patients whose BMI <23 kg/m² and heart rates <65 bpm selected from 291 patients diagnosed of suspected CHD at our institution from September 2013 to February 2014 were randomly divided into 3 groups before CCTA, and there were 45 patients in each group. 80 kV , Iodixanol (320 mgI/ml) and sinogram affirmed iterative reconstruction (SAFIRE) were used in A group. 80 kV , Iopamidol (370 mgI/ml) and SAFIRE were used in B group. 100 kV, Iodixanol and filtered back projection (FBP) were used in C group. Two radiologists assessed image quality with 5-piont scale subjectively and double-blind. Independent-Sample Test was used to analyze statistical significance of image quality including signal to noise ratio(SNR) and contrast to noise ratio (CNR) between A and B group or between A and C group. At the same time, Contrast medium dose statistical significance between A and B group and mean effective Radiation dose (ED)statistical significance between A and C were analyzed by the same way.

RESULTS: There were no significant difference of image quality including SNR and CNR of aortic root (AO), left main coronary artery (LM), left anterior descending artery (LAD), circumflex coronary artery (CX) and right coronary artery (RCA) Between A and B group (P = non-significant for all comparison), whereas Iodine in taken of A group decreased 14% (17 600 mg vs 20 350 mg). There were no significant difference of image quality including SNR and CNR of AO, LM, LAD, CX and RCA Between A and C group (P = non-significant for all comparison), whereas mean ED of A group decreased 50% (0.41 ± 0.05 mSv vs 0.79 ± 0.15 mSv).

CONCLUSION: The double low dose application which use High-pitch spiral mode, 80 kV, SAFIRE, and Iodixanol (320 mgI/ml) can be used in those patients whose BMI <23 kg/m² and heart rates <65 bpm to reduce the burden of radiation and contrast medium significantly, without compromising the image quality.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app