Evaluation of a Rapid Immunochromatographic Treponemal Antibody Test Comparing the Treponema Pallidum Particle Agglutination Assay.
Journal of Clinical Laboratory Analysis 2015 September
BACKGROUND: In addition to conventional tests, several methods for detection of treponema-specific antibodies in clinical settings have been recently introduced. We aim to comparatively evaluate a rapid immunochromatographic test (ICT) for Treponema pallidum specific antibody (SD Bioline Syphilis 3.0) and the T. pallidum particle agglutination (TPPA) assay.
METHODS: In all, 132 serum samples from 78 syphilis patients and 54 syphilis-negative controls were analyzed. SD Bioline Syphilis 3.0 test (Standard Diagnostic, Inc., Yongin, Korea) was evaluated and compared to Serodia TPPA assay (Fujirebio, Inc., Tokyo, Japan). All discrepant results between the two assays were repeatedly tested and evaluated by the fluorescent treponemal antibody-absorption (FTA-ABS) assay. Test reproducibility and 95% limit of detection of SD Bioline Syphilis 3.0 were determined across three different lots for seven consecutive days in triplicate. Interference due to autoantibodies and pregnancy was also tested.
RESULTS: Percent agreement between SD Bioline Syphilis 3.0 and TPPA assays was 99.2%. Sensitivity and specificity were 100%, respectively. In TPPA assay, test-to-test, day-to-day, and lot-to-lot variations were not identified until 1:320 titer (eightfold dilutions). There was no interference due to the presence of antinuclear antibodies or samples or pregnancy.
CONCLUSIONS: Percent agreement of SD Syphilis 3.0 and TPPA was very good. Sensitivity and specificity were appropriate for T. pallidum antibody detection. Thus, a rapid ICT could be suitable for syphilis antibody detection.
METHODS: In all, 132 serum samples from 78 syphilis patients and 54 syphilis-negative controls were analyzed. SD Bioline Syphilis 3.0 test (Standard Diagnostic, Inc., Yongin, Korea) was evaluated and compared to Serodia TPPA assay (Fujirebio, Inc., Tokyo, Japan). All discrepant results between the two assays were repeatedly tested and evaluated by the fluorescent treponemal antibody-absorption (FTA-ABS) assay. Test reproducibility and 95% limit of detection of SD Bioline Syphilis 3.0 were determined across three different lots for seven consecutive days in triplicate. Interference due to autoantibodies and pregnancy was also tested.
RESULTS: Percent agreement between SD Bioline Syphilis 3.0 and TPPA assays was 99.2%. Sensitivity and specificity were 100%, respectively. In TPPA assay, test-to-test, day-to-day, and lot-to-lot variations were not identified until 1:320 titer (eightfold dilutions). There was no interference due to the presence of antinuclear antibodies or samples or pregnancy.
CONCLUSIONS: Percent agreement of SD Syphilis 3.0 and TPPA was very good. Sensitivity and specificity were appropriate for T. pallidum antibody detection. Thus, a rapid ICT could be suitable for syphilis antibody detection.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app