Validation of a novel method for measuring intra-abdominal pressure and gastric residual volume in critically ill patients.

BACKGROUND: Gastric residual volume (GRV) can be measured in a variety of ways in critically ill patients, most often, the nasogastric tube is disconnected and the GRV is aspirated via a 60 mL syringe. Bladder pressure (IBP) measurement is the gold standard for intra-abdominal pressure (IAP) estimation. This study will look at the validation of a novel method combining measurement of GRV and estimation of IAP via intra-gastric pressure (IGP).

METHODS: In total 135 paired IAP and 146 paired GRV measurements were performed in 37 mechanically ventilated ICU patients. The IAP was estimated via the bladder (i.e. IBP) using the FoleyManometer and via the stomach (i.e. IGP) with the new device. The GRV was measured with the new device (GRVprototype) and via the classic method (GRVclassic). The devices were provided by Holtech Medical (Charlottenlund, Denmark) and data were retrospectively analysed.

RESULTS: The number of paired measurements in each patient was 4 ± 1. The mean IBP was 10.7 ± 4.1 and mean IGP was 11.6 ± 4.1 mm Hg. Correlation between the IBP and IGP was significant, however moderate (R2 = 0.51). Analysis according to Bland and Altman showed a bias and precision of 0.8 and 2.7 mm Hg respectively, however the limits of agreement (LA) were large and ranged from -4.5 to 6.1 mm Hg. Changes in IGP correlated well with changes in IBP. The median GRVprototype was 80 mL (0-1050) and equal to the median GRVclassic of 80 mL (0-1250). Correlation between the 2 methods was excellent (R2 = 0.89). Analysis according to Bland and Altman showed a bias and precision of -0.8 and 52.3 mL respectively and the LA ranged from -103 to 102 mL. Changes in GRVclassic correlated well with changes in GRVprototype.

CONCLUSIONS: The results of this multicentre pilot study show that GRV can be measured with the new device. Furthermore this allows simultaneous screening for intra-abdominal hypertension with IAP estimation via IGP.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app