Pseudo-malfunction of the Coloplast Titan Inflatable Penile Prosthesis One-Touch Release Pump.
Urology 2014 October
OBJECTIVE: To define and describe a type of pseudo-malfunction of the Coloplast Titan Inflatable Penile Prosthesis (IPP) One-Touch Release (OTR) pump (Coloplast Corp, Minneapolis, MN).
MATERIALS AND METHODS: We retrospectively reviewed a consecutive series of 550 patients with refractory organic erectile dysfunction who were implanted with a Coloplast Titan IPP with the OTR pump during a period of approximately 4 years.
RESULTS: All patients were implanted using standard techniques through an infrapubic or penoscrotal approach. Twenty-nine patients (5.3%) complained that their IPP would not inflate and that the pump bulb felt "hard." Examination revealed that their IPP was working normally; however, the inflate/deflate valve disc had become stuck in the deflate position. Very firm pressure had to be applied to the pump bulb to move the valve disc into the inflate position. Once this was accomplished, the device inflated and deflated normally. Another 14 patients (2.5%) reported this phenomenon to us but were able to apply enough pressure on the pump bulb to rectify it.
CONCLUSION: The inflate/deflate valve disc in the Coloplast Titan OTR pump can occasionally become stuck in the deflate position (7.8% of patients in our experience). Patients may be unable to inflate the device and return for evaluation. In all cases we have encountered, firm pressure on the pump bulb caused the valve to shift into the inflate position, and the device worked properly thereafter. Patients and implanting urologists should be aware of this issue and of the way in which it can be rectified.
MATERIALS AND METHODS: We retrospectively reviewed a consecutive series of 550 patients with refractory organic erectile dysfunction who were implanted with a Coloplast Titan IPP with the OTR pump during a period of approximately 4 years.
RESULTS: All patients were implanted using standard techniques through an infrapubic or penoscrotal approach. Twenty-nine patients (5.3%) complained that their IPP would not inflate and that the pump bulb felt "hard." Examination revealed that their IPP was working normally; however, the inflate/deflate valve disc had become stuck in the deflate position. Very firm pressure had to be applied to the pump bulb to move the valve disc into the inflate position. Once this was accomplished, the device inflated and deflated normally. Another 14 patients (2.5%) reported this phenomenon to us but were able to apply enough pressure on the pump bulb to rectify it.
CONCLUSION: The inflate/deflate valve disc in the Coloplast Titan OTR pump can occasionally become stuck in the deflate position (7.8% of patients in our experience). Patients may be unable to inflate the device and return for evaluation. In all cases we have encountered, firm pressure on the pump bulb caused the valve to shift into the inflate position, and the device worked properly thereafter. Patients and implanting urologists should be aware of this issue and of the way in which it can be rectified.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app