T2 Relaxation time quantitation differs between pulse sequences in articular cartilage.
BACKGROUND: To compare T2 relaxation time measurements between MR pulse sequences at 3 Tesla in agar phantoms and in vivo patellar, femoral, and tibial articular cartilage.
METHODS: T2 relaxation times were quantified in phantoms and knee articular cartilage of eight healthy individuals using a single echo spin echo (SE) as a reference standard and five other pulse sequences: multi-echo SE (MESE), fast SE (2D-FSE), magnetization-prepared spoiled gradient echo (3D-MAPSS), three-dimensional (3D) 3D-FSE with variable refocusing flip angle schedules (3D vfl-FSE), and quantitative double echo steady state (qDESS). Cartilage was manually segmented and average regional T2 relaxation times were obtained for each sequence. A regression analysis was carried out between each sequence and the reference standard, and root-mean-square error (RMSE) was calculated.
RESULTS: Phantom measurements from all sequences demonstrated strong fits (R(2) > 0.8; P < 0.05). For in vivo cartilage measurements, R(2) values, slope, and RMSE were: MESE: 0.25/0.42/5.0 ms, 2D-FSE: 0.64/1.31/9.3 ms, 3D-MAPSS: 0.51/0.66/3.8 ms, 3D vfl-FSE: 0.30/0.414.2 ms, qDESS: 0.60/0.90/4.6 ms.
CONCLUSION: 2D-FSE, qDESS, and 3D-MAPSS demonstrated the best fits with SE measurements as well as the greatest dynamic ranges. The 3D-MAPSS, 3D vfl-FSE, and qDESS demonstrated the closest average measurements to SE. Discrepancies in T2 relaxation time quantitation between sequences suggest that care should be taken when comparing results between studies.
METHODS: T2 relaxation times were quantified in phantoms and knee articular cartilage of eight healthy individuals using a single echo spin echo (SE) as a reference standard and five other pulse sequences: multi-echo SE (MESE), fast SE (2D-FSE), magnetization-prepared spoiled gradient echo (3D-MAPSS), three-dimensional (3D) 3D-FSE with variable refocusing flip angle schedules (3D vfl-FSE), and quantitative double echo steady state (qDESS). Cartilage was manually segmented and average regional T2 relaxation times were obtained for each sequence. A regression analysis was carried out between each sequence and the reference standard, and root-mean-square error (RMSE) was calculated.
RESULTS: Phantom measurements from all sequences demonstrated strong fits (R(2) > 0.8; P < 0.05). For in vivo cartilage measurements, R(2) values, slope, and RMSE were: MESE: 0.25/0.42/5.0 ms, 2D-FSE: 0.64/1.31/9.3 ms, 3D-MAPSS: 0.51/0.66/3.8 ms, 3D vfl-FSE: 0.30/0.414.2 ms, qDESS: 0.60/0.90/4.6 ms.
CONCLUSION: 2D-FSE, qDESS, and 3D-MAPSS demonstrated the best fits with SE measurements as well as the greatest dynamic ranges. The 3D-MAPSS, 3D vfl-FSE, and qDESS demonstrated the closest average measurements to SE. Discrepancies in T2 relaxation time quantitation between sequences suggest that care should be taken when comparing results between studies.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app