A novel flexible hyper-redundant surgical robot: prototype evaluation using a single incision flexible access pelvic application as a clinical exemplar.

INTRODUCTION: The flexible endoscope is increasingly being considered as a surgical tool to enable innovative natural orifice or flexible access techniques. These experiences have exposed unique advantages but also significant challenges. Major current technical drawbacks in this setting relate to uncontrolled flexibility, inaccurate sustained target localization, unreliable navigation and overall platform instability. In striving to address existing technical limitations, this paper introduces a novel flexible hyper-redundant surgical robot and evaluates its clinical potential using a focused clinical application.

METHOD: To assess utility of the device within tight confines of the human pelvis or peritoneal cavity, detailed laboratory workspace analysis experiments were undertaken using a computer-simulated model that incorporated anatomical data obtained via pelvic magnetic resonance images of eight women. Ten participants executed ninety usability and reliability trials on an ex vivo simulator, before the robot was repeatedly trialled in an in vivo porcine model.

RESULTS: The robot demonstrated capability of targeting >90 % of the anatomic region of interest. All 90 user trials were successfully performed without interruption or malfunction. Significant improvements in performance, time and motion were observed between first and last sets of trials (p = 0.001). In vivo feasibility testing affirmed robustness of the device when deployed within the physiological demands of a live scale appropriate model.

CONCLUSION: Technologically advanced flexible operative platforms are needed to fulfil aspirations for an introductory era of flexible access surgery. This prototype is proposed as a potential future platform for robot-assisted flexible endoscopic surgery. Encouraging pre-clinical feasibility results are demonstrated for diagnostic and therapeutic applications within the pelvis.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app