Prospective nationwide fluoroscopic and electrical longitudinal follow-up of recalled Riata defibrillator leads in Denmark.
BACKGROUND: Recalled St. Jude Medical Riata defibrillator leads are prone to insulation failures with externalized conductors (ECs). Longitudinal studies are needed to guide lead management.
OBJECTIVE: The purpose of this study was to describe the dynamic nature of EC and the association with electrical abnormalities and lead extraction outcomes.
METHODS: A nationwide cohort established in 2012 of 295 patients with recalled Riata leads with dwell time 5.1 ± 1.1 years, 34 ECs, and 19 electrical abnormalities were followed until death, lead discontinuation with fluoroscopy, or a new 2013 screening with fluoroscopy and device interrogation.
RESULTS: Fluoroscopic follow-up of 239 patients with normal baseline fluoroscopy revealed incident overt EC in 8 leads and borderline EC in 2 leads after 1.1 ± 0.2 years, with an incidence rate of 3.7 per 100 person-years (95% confidence interval 2.0-6.9). Fluoroscopic follow-up in 27 patients with baseline EC showed an increase in EC length of 4 ± 1 mm (P <.001) after 1.1 ± 0.3 years. Electrical follow-up in 276 patients with normal baseline electrical function demonstrated 20 incident electrical abnormalities after 1.0 ± 0.3 years, with an incidence rate of 7.1 per 100 person-years (95% confidence interval 4.6-11.0). This rate was significantly higher in leads with baseline EC, with an adjusted incidence rate ratio of 4.4 (95% confidence interval 1.7-11.5, P = .002). In 15 extractions, all leads were removed, with 2 major complications.
CONCLUSION: The development of EC is a dynamic process despite long lead dwell time. ECs are associated with a higher risk of electrical abnormalities. Therefore, lead replacement should be considered, especially in patients with a long life expectancy.
OBJECTIVE: The purpose of this study was to describe the dynamic nature of EC and the association with electrical abnormalities and lead extraction outcomes.
METHODS: A nationwide cohort established in 2012 of 295 patients with recalled Riata leads with dwell time 5.1 ± 1.1 years, 34 ECs, and 19 electrical abnormalities were followed until death, lead discontinuation with fluoroscopy, or a new 2013 screening with fluoroscopy and device interrogation.
RESULTS: Fluoroscopic follow-up of 239 patients with normal baseline fluoroscopy revealed incident overt EC in 8 leads and borderline EC in 2 leads after 1.1 ± 0.2 years, with an incidence rate of 3.7 per 100 person-years (95% confidence interval 2.0-6.9). Fluoroscopic follow-up in 27 patients with baseline EC showed an increase in EC length of 4 ± 1 mm (P <.001) after 1.1 ± 0.3 years. Electrical follow-up in 276 patients with normal baseline electrical function demonstrated 20 incident electrical abnormalities after 1.0 ± 0.3 years, with an incidence rate of 7.1 per 100 person-years (95% confidence interval 4.6-11.0). This rate was significantly higher in leads with baseline EC, with an adjusted incidence rate ratio of 4.4 (95% confidence interval 1.7-11.5, P = .002). In 15 extractions, all leads were removed, with 2 major complications.
CONCLUSION: The development of EC is a dynamic process despite long lead dwell time. ECs are associated with a higher risk of electrical abnormalities. Therefore, lead replacement should be considered, especially in patients with a long life expectancy.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app