Pulmonary vein stenosis after catheter ablation: electroporation versus radiofrequency.
Circulation. Arrhythmia and Electrophysiology 2014 August
BACKGROUND: Radiofrequency ablation inside pulmonary vein (PV) ostia can cause PV stenosis. A novel alternative method of ablation is irreversible electroporation, but the long-term response of PVs to electroporation ablation is unknown.
METHODS AND RESULTS: In ten 6-month-old pigs (60-75 kg), the response of PVs to circular electroporation and radiofrequency ablation was compared. Ten consecutive, nonarcing, electroporation applications of 200 J were delivered 5 to 10 mm inside 1 of the 2 main PVs, using a custom-deflectable, 18-mm circular decapolar catheter. Inside the other PV, circular radiofrequency ablation was performed using 30 W radiofrequency applications via an irrigated 4-mm ablation catheter. PV angiograms were made before ablation, immediately after ablation, and after 3-month survival. PV diameters and heart size were measured. With electroporation ablation, PV ostial diameter decreased 11±10% directly after ablation, but had increased 19±11% after 3 months. With radiofrequency ablation, PV ostial diameter decreased 23±15% directly after ablation and remained 7±17% smaller after 3 months compared with preablation diameter despite a 21±7% increase in heart size during aging from 6 to 9 months.
CONCLUSIONS: In this porcine model, multiple circumferential 200-J electroporation applications inside the PV ostia do not affect PV diameter at 3-month follow-up. Radiofrequency ablation inside PV ostia causes considerable PV stenosis directly after ablation, which persists after 3 months.
METHODS AND RESULTS: In ten 6-month-old pigs (60-75 kg), the response of PVs to circular electroporation and radiofrequency ablation was compared. Ten consecutive, nonarcing, electroporation applications of 200 J were delivered 5 to 10 mm inside 1 of the 2 main PVs, using a custom-deflectable, 18-mm circular decapolar catheter. Inside the other PV, circular radiofrequency ablation was performed using 30 W radiofrequency applications via an irrigated 4-mm ablation catheter. PV angiograms were made before ablation, immediately after ablation, and after 3-month survival. PV diameters and heart size were measured. With electroporation ablation, PV ostial diameter decreased 11±10% directly after ablation, but had increased 19±11% after 3 months. With radiofrequency ablation, PV ostial diameter decreased 23±15% directly after ablation and remained 7±17% smaller after 3 months compared with preablation diameter despite a 21±7% increase in heart size during aging from 6 to 9 months.
CONCLUSIONS: In this porcine model, multiple circumferential 200-J electroporation applications inside the PV ostia do not affect PV diameter at 3-month follow-up. Radiofrequency ablation inside PV ostia causes considerable PV stenosis directly after ablation, which persists after 3 months.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app